Task-based image quality assessment of an intraoperative CBCT for spine surgery compared with conventional CT
- PMID: 38986263
- DOI: 10.1016/j.ejmp.2024.103426
Task-based image quality assessment of an intraoperative CBCT for spine surgery compared with conventional CT
Abstract
Purpose: To analyze the image quality of a novel, state-of-the art platform for CBCT image-guided spine surgery, focusing particularly on the dose-effectiveness compared with conventional CT (the gold standard for postoperative assessment).
Methods: The ClarifEye platform (Philips Healthcare) with integrated augmented-reality surgical navigation, has been compared with a GE Revolution CT (GE Healthcare). The 3D spatial resolution (TTF) and noise (NPS) were evaluated considering relevant feature contrasts (200-900 HU) and background noise for differently sized patients (200-300 mm water-equivalent diameter). These measures were used to determine the noise equivalent quanta (NEQ) and observer model detectability.
Results: The CBCT system exhibited a linear response with 50% TTF at 5.7 cycles/cm (10% TTF at 9.2 cycles/cm), and the axial noise power peaking at about 3.6 cycles/cm (average frequency of 4.1 cycles/cm). The noise magnitude and texture differed markedly compared to iteratively reconstructed CT images (GE ASiR-V). The CBCT system had 26% lower detectability for a high-frequency task (related to edge detection) compared with CT images reconstructed using the Bone kernel combined with ASiR-V 50%. Likewise, it had 18% lower detectability for low- and mid-frequency tasks compared with CT images reconstructed using the Standard kernel. This difference translates to 50%-80% higher CBCT imaging doses required to match the CT image quality.
Conclusions: The ClarifEye platform demonstrates intraoperative CBCT-imaging capabilities that under certain circumstances are comparable with conventional CT. However, due to limited dose-effectiveness, a trade-off between timeliness and radiation exposure must be considered if end-of-procedure CBCT is to replace postoperative CT.
Keywords: Intraoperative CBCT; Noise-equivalent quanta; Observer model detectability; Postoperative CT.
Copyright © 2024 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.
Conflict of interest statement
Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Similar articles
-
Mobile C-arm cone-beam CT for guidance of spine surgery: image quality, radiation dose, and integration with interventional guidance.Med Phys. 2011 Aug;38(8):4563-74. doi: 10.1118/1.3597566. Med Phys. 2011. PMID: 21928628 Free PMC article.
-
Radiation dose and image quality comparison during spine surgery with two different, intraoperative 3D imaging navigation systems.J Appl Clin Med Phys. 2019 Feb;20(2):136-145. doi: 10.1002/acm2.12534. Epub 2019 Jan 24. J Appl Clin Med Phys. 2019. PMID: 30677233 Free PMC article.
-
Noise and spatial resolution properties of a commercially available deep learning-based CT reconstruction algorithm.Med Phys. 2020 Sep;47(9):3961-3971. doi: 10.1002/mp.14319. Epub 2020 Jul 6. Med Phys. 2020. PMID: 32506661
-
A mobile isocentric C-arm for intraoperative cone-beam CT: Technical assessment of dose and 3D imaging performance.Med Phys. 2020 Mar;47(3):958-974. doi: 10.1002/mp.13983. Epub 2020 Jan 6. Med Phys. 2020. PMID: 31863480 Free PMC article.
-
A quality assurance framework for the fully automated and objective evaluation of image quality in cone-beam computed tomography.Med Phys. 2014 Mar;41(3):031901. doi: 10.1118/1.4863507. Med Phys. 2014. PMID: 24593719
Cited by
-
Gaussian Function Model for Task-Specific Evaluation in Medical Imaging: A Theoretical Investigation.J Imaging Inform Med. 2025 Apr 24. doi: 10.1007/s10278-025-01511-9. Online ahead of print. J Imaging Inform Med. 2025. PMID: 40274693
-
Image quality assessment in spine surgery: a comparison of intraoperative CBCT and postoperative MDCT.Acta Neurochir (Wien). 2025 Mar 31;167(1):94. doi: 10.1007/s00701-025-06503-w. Acta Neurochir (Wien). 2025. PMID: 40164732 Free PMC article.
-
Image Quality Comparison of Three 3D Mobile X-Ray Imaging Guidance Devices Used in Spine Surgery: A Phantom Study.Sensors (Basel). 2024 Oct 26;24(21):6883. doi: 10.3390/s24216883. Sensors (Basel). 2024. PMID: 39517780 Free PMC article.
-
Comparison of Effective Dose From Intraoperative 3D Imaging Using Mobile and Fixed C-Arms Versus Computed Tomography When Evaluating the Lumbar Spine.Global Spine J. 2025 Apr 17:21925682251334986. doi: 10.1177/21925682251334986. Online ahead of print. Global Spine J. 2025. PMID: 40245478 Free PMC article.
-
[Advances in low-dose cone-beam computed tomography image reconstruction methods based on deep learning].Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2025 Jun 25;42(3):635-642. doi: 10.7507/1001-5515.202409021. Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2025. PMID: 40566788 Free PMC article. Review. Chinese.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources