Patient radiation risk reduction by controlling the tube start angle in single and dual source spiral CT scans: A simulation study
- PMID: 38991121
- DOI: 10.1002/mp.17290
Patient radiation risk reduction by controlling the tube start angle in single and dual source spiral CT scans: A simulation study
Abstract
Background: Organ doses in spiral CT scans depend on the tube start angle.
Purpose: To determine the effective dose in single source CT (SSCT) and dual source CT (DSCT) scans as a function of tube start angle and spiral pitch value to identify the dose reduction potential by selecting the optimal start angle.
Methods: Using Monte Carlo simulations, dose values for different tube positions with an angular increment of and a longitudinal increment of were simulated over a range of with collimations of , , and . The simulations were performed for the thorax region of six adult patients based on clinical CT data. From the resulting dose distributions, organ doses and effective dose were determined as a function of tube angle and longitudinal position. Using these per-view dose data, the individual organ doses, as well as the total effective dose, were determined for spiral scans with and without tube current modulation (TCM) with pitch values ranging from 0.5 to 1.5 for SSCT and up to 3.0 for DSCT. The dose of the best and worst tube start angle in terms of dose was determined and compared to the mean dose over all tube start angles.
Results: With increasing pitch and collimation, the dose variations from the effective dose averaged over all start angles increase. While for a collimation of , the variations from the mean dose value stay below for SSCT, we find that for a spiral scan with a pitch of 3.0 for DSCT with TCM and collimation of , the dose for the best starting angle is on average lower than the mean value and lower than the maximum value.
Conclusions: Variation of the tube start angle in spiral scans exhibits substantial differences in radiation dose especially for high pitch values and for high collimations. Therefore, we suggest to control the tube start angle to minimize patient risk.
Keywords: computed tomography; radiation risk; spiral CT; tube start angle.
© 2024 The Author(s). Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.
Similar articles
-
Reducing radiation dose to selected organs by selecting the tube start angle in MDCT helical scans: a Monte Carlo based study.Med Phys. 2009 Dec;36(12):5654-64. doi: 10.1118/1.3259773. Med Phys. 2009. PMID: 20095278 Free PMC article.
-
Dual-source spiral CT with pitch up to 3.2 and 75 ms temporal resolution: image reconstruction and assessment of image quality.Med Phys. 2009 Dec;36(12):5641-53. doi: 10.1118/1.3259739. Med Phys. 2009. PMID: 20095277
-
Very low-dose coronary artery calcium scanning with high-pitch spiral acquisition mode: comparison between 120-kV and 100-kV tube voltage protocols.J Cardiovasc Comput Tomogr. 2013 Jan-Feb;7(1):32-8. doi: 10.1016/j.jcct.2012.11.004. Epub 2012 Dec 1. J Cardiovasc Comput Tomogr. 2013. PMID: 23333186
-
Radiation dose considerations by intra-individual Monte Carlo simulations in dual source spiral coronary computed tomography angiography with electrocardiogram-triggered tube current modulation and adaptive pitch.Eur Radiol. 2012 Mar;22(3):569-78. doi: 10.1007/s00330-011-2300-6. Epub 2011 Oct 9. Eur Radiol. 2012. PMID: 21984448
-
Intra-individual diagnostic image quality and organ-specific-radiation dose comparison between spiral cCT with iterative image reconstruction and z-axis automated tube current modulation and sequential cCT.Eur J Radiol Open. 2016 Jul 26;3:182-90. doi: 10.1016/j.ejro.2016.05.006. eCollection 2016. Eur J Radiol Open. 2016. PMID: 27504476 Free PMC article. Review.
References
REFERENCES
-
- Euler A, Szucs‐Farkas Z, Falkowski AL, et al. Organ‐based tube current modulation in a clinical context: dose reduction may be largely overestimated in breast tissue. Eur Radiol. 2016;26:2656‐2662.
-
- Ota J, Yokota H, Takishima H, et al. Breast exposure reduction using organ‐effective modulation on chest CT in Asian women. Eur J Radiol. 2019;119:108651.
-
- Papadakis AE, Damilakis J. Evaluation of an organ‐based tube current modulation tool in pediatric CT examinations. Eur Radiol. 2020;30:5728‐5737.
-
- Klein L, Liu C, Steidel J, et al. Patient‐specific radiation risk‐based tube current modulation for diagnostic CT. Med Phys. 2022;49:4391‐4403.
-
- Zhang D, Savandi AS, Demarco JJ, et al. Variability of surface and center position radiation dose in MDCT: Monte Carlo simulations using CTDI and anthropomorphic phantoms. Med Phys. 2009;36:1025‐1038.
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources