Randomized Controlled Trial
. 2024 Sep;10(9):2088-2096.
doi: 10.1016/j.jacep.2024.04.034. Epub 2024 Jul 10.

Effects of Implantable Cardioverter-Defibrillator Leads on the Tricuspid Valve and Right Ventricle: A Randomized Trial

Affiliations
Free article
Randomized Controlled Trial

Effects of Implantable Cardioverter-Defibrillator Leads on the Tricuspid Valve and Right Ventricle: A Randomized Trial

Darryl P Leong et al. JACC Clin Electrophysiol. 2024 Sep.
Free article

Abstract

Background: There are no randomized data to inform the extent to which transvenous cardiac leads cause tricuspid regurgitation (TR).

Objectives: This study sought to determine the effect of a transvenous implantable cardioverter-defibrillator (TV-ICD) on TR severity, and secondarily, on right ventricular (RV) size and function.

Methods: We evaluated TR severity before and 6 months after implantable cardioverter-defibrillator insertion in a post hoc analysis of adults randomized to receive a transvenous (n = 252) or subcutaneous implantable cardioverter-defibrillator (S-ICD) (n = 251) device. TR and RV size and systolic function were assessed by echocardiographic images analyzed in a core laboratory.

Results: At baseline, at least mild TR was present in 30% of individuals. At 6 months, the proportion of participants with any TR in the TV-ICD group was 42% vs 19% in the S-ICD group (P < 0.001). The proportion with moderate or severe TR was 7% in the TV-ICD group vs 2% in the S-ICD group (P = 0.021). At 6 months, the OR of at least 1 grade worsening of TR in the TV-ICD group as compared with the S-ICD group was 7.2 (95% CI: 3.3-15.8; P < 0.001). There were no differences between groups with respect to RV size or systolic function.

Conclusions: Six months following TV-ICD insertion, there was a 7-fold increase in the risk of at least 1 grade worsening of TR, with 7% of individuals having TR that was moderate or severe. There was no detectable difference in RV size or function; however, longer follow-up is needed.

Keywords: cardiovascular implantable electronic device; implantable cardioverter-defibrillator; lead; right ventricular; tricuspid regurgitation.

PubMed Disclaimer

Conflict of interest statement

Funding Support and Author Disclosures This study was funded by Boston Scientific through an unrestricted research grant. Boston Scientific did not have any input into the preparation of this manuscript. Dr Leong has received consultancy fees from Boston Scientific. Dr Mondésert has received consultancy and speaking fees from Bostin Scientific, Abbott, Medtronic, and Biotronik. Dr Joza has received research support from Medtronic Inc; consulting fees from Boston Scientific; and honoraria from Biosense Webster Canada. Dr Healey has received research support and speaking fees from BMS/Pfizer, Boston Scientific, and Medtronic; speaking fees from Servier; and consultancy fees from Bayer and Boston Scientific. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose.