Multicenter Study
. 2024 Nov;80(5):1317-1325.e2.
doi: 10.1016/j.jvs.2024.07.086. Epub 2024 Jul 26.

Proximal sealing in the aortic arch for inner curve disease using the custom Relay scalloped and fenestrated stent graft

Collaborators, Affiliations
Free article
Multicenter Study

Proximal sealing in the aortic arch for inner curve disease using the custom Relay scalloped and fenestrated stent graft

Simona Sica et al. J Vasc Surg. 2024 Nov.
Free article

Abstract

Objective: This study aimed to analyze early and midterm results of custom-made proximal scallop and fenestrated stent grafts for thoracic endovascular aortic repair (TEVAR) with a proximal landing zone (PLZ) in the aortic arch.

Methods: All consecutive patients treated with the custom made proximal scalloped and fenestrated Relay stent grafts (Terumo Aortic Bolton Medical Inc.) in 10 Italian centers between January 2014 and December 2022 were included. The primary end points were technical success, incidence of intraoperative major adverse events, deployment accuracy, and rate of early neurological complications, endoleaks (ELs) and retrograde aortic dissection.

Results: During the study period, 49 patients received TEVAR with Relay custom-made endograft in Italy were enrolled. The median patient age was 70.1 years (interquartile range, 23-86 years) and 65.3% were male. The indication for treatment was atherosclerotic aneurysms in 59.2% of cases and penetrating aortic ulcer in 22.4%. The endograft configuration was proximal fenestration in 55.1% and scallop in 44.9%. The proximal landing zone was zone 0 in 25 cases (51%), zone 1 in 14 cases (28.6%), and zone 2 in 10 cases (20.4%). The supra-aortic debranching procedures were 38 (77.5%). Technical success was 97.9% (48/49) owing to one case (2.0%) of inaccurate deployment. Intraoperatively, one (2.0%) type Ia and one (2.0%) type III EL were detected. There were no cases of in-hospital mortality, major adverse events, or retrograde dissection. Three minor strokes (6.1%) (National Institutes of Health Stroke Scale score of ≤4) were observed. At a mean follow-up time of 36.3 ± 21.3 months the rate of types I to III ELs and reintervention was 4.1%, respectively. Four patients (8.2%) died during the follow-up period, one (2.1%) from abdominal aortic rupture and three (6.1%) from nonaortic causes.

Conclusions: Our early and midterm outcomes suggest that scalloped and fenestrated TEVAR may provide an acceptable alternative treatment option for aortic arch pathologies. Large-scale studies are needed to assess the long-term durability of this technique.

Keywords: Aortic arch; Custom-made; Fenestrated; Multicenter study; Scallop; Stent graft; TEVAR.

PubMed Disclaimer

Conflict of interest statement

Disclosures None.

Publication types

MeSH terms