Abstract
Diabetes is a chronic disease with significant complications, necessitating regular treatment and checkups, which can be costly and time-consuming for patients. To address this, we developed the Sliding Microneedle (MN)-Lateral flow immunoassay strip (LFIAs) device that combines the advantages of MNs and LFIAs to detect IL-6, an independent biomarker for diabetes complications. This device offers rapid and highly sensitive detection of IL-6 by extracting interstitial fluid (ISF) through MNs and transferring it to LFIAs. The stainless MN, embedded in the 3D-printed Sliding MN-LFIAs device, was inserted into the skin at a 20° angle, minimizing blood contamination risk. With a filter paper attached to the MN surface, the device collected 4.65 ± 0.05 μL of ISF containing IL-6 within 90 s. The ISF was then transferred to the LFIAs using a running buffer. After a 15-min reaction, silver enhancement (SE) treatment was applied, allowing for the highly sensitive and specific detection of IL-6 at 102 pg/mL concentrations. The Sliding MN-LFIAs device successfully distinguished between normal and diabetic rat models, demonstrating its potential as an effective tool for detecting diabetes complications quickly and affordably.
Keywords: Diabetes complications; Interleukin-6; Interstitial fluid; Lateral flow immunoassay strip; Microneedle.
Copyright © 2024 Elsevier B.V. All rights reserved.
Conflict of interest statement
Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources