Axillary and subclavian venous spasm during pacemaker implantation - A case report and literature review
- PMID: 39139620
- PMCID: PMC11319219
- DOI: 10.1002/ccr3.9309
Axillary and subclavian venous spasm during pacemaker implantation - A case report and literature review
Abstract
Key clinical message: Venous spasm is an important reason for complicated or failed implantations of cardiac implantable electronic devices. Prevention or risk reduction of venous spasm during cardiac implantable electronic device implantation may be achieved by ultrasound or fluoroscopic imaging prior to puncture, cephalic vein cut-down, sufficient pre- and perioperative hydration, nitroglycerin injection and effective sedation, and analgesia.
Abstract: This case report with literature review focuses on venous spasm as a potential cause for complicated implantations of cardiac implantable electronic devices. The case report is clinically relevant as it describes a progressive spasm affecting the axillary and the subclavian vein. A 66-year-old female complained of symptomatic atrial fibrillation (AF) and atypical atrial flutter despite interventional and medical treatment. As an ultimate treatment, she was scheduled for pacemaker implantation and atrioventricular node ablation. Several puncture attempts of the axillary vein failed. Despite venous blood aspiration, no guidewires could be advanced into the axillary vein. We performed a first venogram revealing significant spasm of the axillary vein. Another failed venous puncture occurred after change of access site to the subclavian vein. A second venogram displayed progression of the spasm, now affecting both the axillary and the subclavian veins. Normal saline perfusion was administered as well as intravenous isosorbide. Unfortunately, a repeated venogram after 15 min waiting time showed persistence of the spasm, still affecting both veins. The procedure was discontinued as the patient became uncomfortable. Venous spasm is an important reason for complicated or failed implantations of cardiac implantable electronic devices. Commonly used medical prevention and treatment are intravenous fluids and nitroglycerin. Prevention or risk reduction of venous spasm during cardiac implantable electronic device implantation may be achieved by ultrasound or fluoroscopic imaging prior to puncture, cephalic vein cut-down, sufficient pre- and perioperative hydration, nitroglycerin injection and effective sedation and analgesia.
Keywords: axillary vein; cardiac implantable electronic device; implantation; pacemaker; subclavian vein; venous spasm.
© 2024 The Author(s). Clinical Case Reports published by John Wiley & Sons Ltd.
Conflict of interest statement
The authors declare no conflicts of interest.
Figures

Similar articles
-
Literature Review of Venous Spasm During Pacemaker Implantation.Cureus. 2025 Jul 24;17(7):e88682. doi: 10.7759/cureus.88682. eCollection 2025 Jul. Cureus. 2025. PMID: 40709021 Free PMC article. Review.
-
Cephalic vein cut-down technique for severe venous spasm following axillary vein puncture at pacemaker implantation.J Cardiol Cases. 2022 Jun 1;26(4):245-247. doi: 10.1016/j.jccase.2022.05.005. eCollection 2022 Oct. J Cardiol Cases. 2022. PMID: 36187320 Free PMC article.
-
Efficacy and safety of nitroglycerin for preventing venous spasm during contrast-guided axillary vein puncture for pacemaker or defibrillator leads implantation.Europace. 2013 Apr;15(4):566-9. doi: 10.1093/europace/eus233. Epub 2012 Jul 29. Europace. 2013. PMID: 22848074 Clinical Trial.
-
Axillary Vein Spasm During Permanent Pacemaker Implantation.J Invasive Cardiol. 2022 Nov;34(11):E824. doi: 10.25270/jic/22.00215. J Invasive Cardiol. 2022. PMID: 36318464
-
How to perform extrathoracic venous access for cardiac implantable electronic device placement: Detailed description of techniques.Heart Rhythm. 2022 Jul;19(7):1184-1191. doi: 10.1016/j.hrthm.2022.02.020. Epub 2022 Feb 26. Heart Rhythm. 2022. PMID: 35231611 Review.
Cited by
-
Literature Review of Venous Spasm During Pacemaker Implantation.Cureus. 2025 Jul 24;17(7):e88682. doi: 10.7759/cureus.88682. eCollection 2025 Jul. Cureus. 2025. PMID: 40709021 Free PMC article. Review.
References
-
- Chen C‐W, Lin C‐L, Lin T‐K, Lin C‐D. A simple and effective regimen for prevention of radial artery spasm during coronary catheterization. Cardiology. 2005;105:43‐47. - PubMed
-
- Dharma S, Shah S, Radadiya R, Vyas C, Pancholy S, Patel T. Nitroglycerin plus diltiazem versus nitroglycerin alone for spasm prophylaxis with transradial approach. J Invasive Cardiol. 2012;24:122‐125. - PubMed
-
- Lanza GA, Careri G, Crea F. Mechanisms of coronary artery spasm. Circulation. 2011;124:1774‐1782. - PubMed
-
- Herrick AL. Pathogenesis of Raynaud's phenomenon. Rheumatology. 2005;44:587‐596. - PubMed
-
- Duan X, Ling F, Shen Y, Yang J, Xu H. Venous spasm during contrast‐guided axillary vein puncture for pacemaker or defibrillator lead implantation. Eurospace. 2012;14:1008‐1011. - PubMed