Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2024 Jul;11(3):035007.
doi: 10.1117/1.NPh.11.3.035007. Epub 2024 Aug 13.

Optical constraints on two-photon voltage imaging

Affiliations

Optical constraints on two-photon voltage imaging

F Phil Brooks 3rd et al. Neurophotonics. 2024 Jul.

Abstract

Significance: Genetically encoded voltage indicators (GEVIs) are a valuable tool for studying neural circuits in vivo, but the relative merits and limitations of one-photon (1P) versus two-photon (2P) voltage imaging are not well characterized.

Aim: We consider the optical and biophysical constraints particular to 1P and 2P voltage imaging and compare the imaging properties of commonly used GEVIs under 1P and 2P excitation.

Approach: We measure the brightness and voltage sensitivity of voltage indicators from commonly used classes under 1P and 2P illumination. We also measure the decrease in fluorescence as a function of depth in the mouse brain. We develop a simple model of the number of measurable cells as a function of reporter properties, imaging parameters, and desired signal-to-noise ratio (SNR). We then discuss how the performance of voltage imaging would be affected by sensor improvements and by recently introduced advanced imaging modalities.

Results: Compared with 1P excitation, 2P excitation requires 10 4 -fold more illumination power per cell to produce similar photon count rates. For voltage imaging with JEDI-2P in the mouse cortex with a target SNR of 10 (spike height to baseline shot noise), a measurement bandwidth of 1 kHz, a thermally limited laser power of 200 mW, and an imaging depth of > 300 μ m , 2P voltage imaging using an 80-MHz source can record from no more than 12 neurons simultaneously.

Conclusions: Due to the stringent photon-count requirements of voltage imaging and the modest voltage sensitivity of existing reporters, 2P voltage imaging in vivo faces a stringent tradeoff between shot noise and tissue photodamage. 2P imaging of hundreds of neurons with high SNR at a depth of > 300 μ m will require either major improvements in 2P GEVIs or qualitatively new approaches to imaging.

Keywords: shot noise; two-photon; voltage imaging.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Comparison of 1P and 2P brightness and sensitivity of fluorescent voltage indicators. (a) Diagram of the experiment. HEK cells were sequentially illuminated with wide-field 1P light in steps of increasing intensity and then by spiral scan 2P steps of increasing intensity. (b) Example HEK cell expressing the GEVI ASAP3. The 2P spiral scan pattern is shown in red, and the analysis ROI is shown in green. Scale bars=5  μm. (c) Example single-trial data for a cell expressing ASAP3. (d) Top: fluorescence in panel (c) as a function of 1P intensity with a linear fit. Bottom: fluorescence as a function of 2P power with quadratic fit. (e) Log–log plot of count rate versus optical power on the cell for seven voltage indicators. 1P data, filled symbols; 2P data, empty symbols. Error bars are standard error of the mean from at least n=8 cells. The excitation wavelengths used for 1P (2P) excitation of each of the reporters were ASAP3 488 nm (930 nm), BeRST1 635 nm (850 nm), JEDI-2P 488 nm (930 nm), QuasAr6a 635 nm (900 nm), Voltron2525 488 nm (930 nm), Voltron2585 594 nm (1100 nm), and Voltron2669 635 nm (1220 nm). A horizontal line is shown at 1.5×107  counts/s, equivalent on our camera to 107 impinging photons/s. (f) Whole-cell patch clamp protocol for measuring voltage sensitivity under 1P and 2P excitation. (g) Average voltage responses of JEDI-2P and Voltron2525 under 1P and 2P illumination. (h) Ratio of voltage contrast under 2P versus 1P illumination for JEDI-2P and Voltron2525, n=5 cells per construct.
Fig. 2
Fig. 2
Depth-dependent 1P and 2P signals in the brain. (a) Experimental protocol. In a mouse expressing JEDI-2P, raster-scanned 2P (λ=920  nm) and DMD-patterned 1P imaging (λ=488  nm) were alternately applied to neurons at different depths. 1P illumination patterned to cell-free regions was used to estimate the background signal. (b) Example 1P and 2P images of cells at three depths. Scale bars=5  μm. (c) Estimated signal-to-background ratio for n=43 neurons under patterned 1P illumination. (d) Mean count rate from the membranes of n=43 neurons under 2P illumination. The color indicates excitation power. The inset shows the cell at 473  μm depth (boxed on the graph). Scale bar=5  μm.
Fig. 3
Fig. 3
Scaling of 2P voltage measurements with depth and with GEVI properties. (a) Predicted number of simultaneously measurable cells as a function of depth, based on brightness derived from HEK cells expressing JEDI-2P (Fig. 1). We assumed a spike contrast of β=0.2; target SNRs of 3 (green dotted), 5 (blue dashed), or 10 (red solid) in an integration time τ=1  ms; a total power P0=200  mW, targeting fraction ϕ=1, scattering length le=112  μm, and a detector with perfect quantum efficiency. (b) Predicted number of simultaneously measurable cells at SNR=10 and 200 mW power for each experimentally measured single-cell count rate reported in Fig. 2(d). (c) Number of simultaneously measurable cells under 2P illumination at a depth of 500  μm, assuming brightness and contrast improvements of future GEVIs, target SNR of 10, and all other parameters as in panel (a).
Fig. 4
Fig. 4
Effect of reporter kinetics on signal. (a) Response of a reporter to a 1-ms voltage pulse. An exponential rise with time constant τon reaches a maximum of β followed by an exponential decay with time constant τoff. The signal comprises the total area under both response phases (cross-hatched). (b) Area under the curve in panel (a) as a function of τon and τoff, keeping constant pulse duration (1 ms) and steady-state voltage sensitivity (βss=0.2). Increasing τoff allows longer integration time, whereas increasing τon truncates the response. (c) Effects on SNR of changing τon or τoff while keeping the other fixed. The fixed parameter is shown in the legend, and variable 1 is indicated by the x-axis. At large τon, SNR1/τon. At large τoff, SNRτoff1/2.
Fig. 5
Fig. 5
Scaling of fluorescence with numerical aperture, sample geometry, and excitation modality. (a) Geometry of a Gaussian beam, showing the width (W01/NA) and waist (b1/NA2). (b) Scaling of total 2P fluorescence as a function of excitation NAe for different sample geometries. All slender dimensions are assumed to be W0, and all extended dimensions are assumed to be b. In a planar raster scan, the fraction of time that a subwavelength structure is excited, ϕ, depends on the focus width and hence the NAe. In cases (iii), (v), and (vi), we assume that the object is perfectly in focus, i.e., in the axial plane where focus size is minimum. The collection efficiency for all geometries depends on the collection solid angle NAc2. To calculate the total signal for targeted illumination, multiply the first and third lines; for a raster scan, multiply all three lines. (c) Total fluorescence evoked by the intersection of a laser focus and a spherical membrane, 10  μm diameter. We compared 1P and 2P excitations with equal NAc and NAe, with powers adjusted to match per-molecule excitation rates at the focus at NA = 1.2. The much smaller 2P focal volume led to a 5.3-fold smaller maximum fluorescence at the highest NA (and even greater discrepancy at lower NA) and a 3-fold greater sensitivity to misalignment compared with 1P excitation.
Fig. 6
Fig. 6
Optimal temporal and spatial splitting. (a) At fixed total power, the per-spot pulse energy is inversely proportional to the effective repetition rate, feff=f×Nspots. SNR is increased by increasing focal pulse energy up to the threshold (1-nJ horizontal line shown). Therefore, the optimal effective repetition rate lies at the intersection of the iso-power line with the threshold (circled in red). At a nonzero depth (dotted lines, le = attenuation length), a lower repetition rate is needed to produce the same focal pulse energy. (b) For a single diffraction-limited focal spot, a total power of 200 mW, and a pulse energy threshold of 1 nJ, the optimal feff goes beneath 80 MHz (horizonal line) at 100  μm depth. Decreasing the threshold focal energy (e.g., to 0.5 nJ) increases the optimal repetition rate at a given depth. (c) For a fixed feff, the focal pulse energy decays exponentially with depth. At feff=250  MHz, a 1-nJ pulse is not achievable at any depth. At feff=40  MHz, the pulse energy crosses the 1 nJ threshold at 200  μm depth.

Update of

Similar articles

Cited by

References

    1. Abdelfattah A. S., et al. , “Bright and photostable chemigenetic indicators for extended in vivo voltage imaging,” Science 365, 699–704 (2019).SCIEAS10.1126/science.aav6416 - DOI - PubMed
    1. Kannan M., et al. , “Dual-polarity voltage imaging of the concurrent dynamics of multiple neuron types,” Science 378, eabm8797 (2022).SCIEAS10.1126/science.abm8797 - DOI - PMC - PubMed
    1. Weber T. D., et al. , “High-speed multiplane confocal microscopy for voltage imaging in densely labeled neuronal populations,” Nat. Neurosci. 26, 1642–1650 (2023).NANEFN10.1038/s41593-023-01408-2 - DOI - PMC - PubMed
    1. Villette V., et al. , “Ultrafast two-photon imaging of a high-gain voltage indicator in awake behaving mice,” Cell 179, 1590–1608.e23 (2019).CELLB510.1016/j.cell.2019.11.004 - DOI - PMC - PubMed
    1. Sims R. R., et al. , “Scanless two-photon voltage imaging,” Nat. Commun. 15, 5095 (2023).NCAOBW10.1038/s41467-024-49192-2 - DOI - PMC - PubMed