Review
. 2025 Jan;168(1):118-125.
doi: 10.1002/ijgo.15860. Epub 2024 Aug 22.

Good practice with fluid management in operative hysteroscopy

Affiliations
Review

Good practice with fluid management in operative hysteroscopy

George A Vilos et al. Int J Gynaecol Obstet. 2025 Jan.

Abstract

Hysteroscopic surgery requires a balance of continuous controlled irrigation and aspiration to distend the endometrial cavity to a degree that provides the clear and stable visual environment necessary for diagnostic and therapeutic procedures. Whereas the preferred distending solution should be isotonic and isonatremic, radiofrequency (RF) electrosurgery with monopolar instrumentation can only be performed with non-ionic (hyponatremic) solutions. Absorption of as little as 500 mL and certainly more than 1000 mL of non-ionic solutions can result in fluid overload and/or dilutional hyponatremia with potentially serious adverse effects under certain conditions and patient characteristics. Both hysteroscopic RF electrosurgery with bipolar instrumentation and electro-mechanical morcellation and aspiration systems use isotonic and isonatremic solutions. Depending on the clinical context, absorption of more than 1500 mL of isonatremic solutions can also result in serious adverse effects. Automated fluid management systems are preferred and recommended, and surgeons should aim to maintain the maximum allowable intravasation of distending media below 1000 and 1500 mL for non-ionic and ionic fluids, respectively.

Keywords: distending media; fluid overload; hyponatremia; hysteroscopic morcellators; hysteroscopic surgery; hysteroscopy.

PubMed Disclaimer

References

REFERENCES

    1. Baker VL, Adamson GD. Minimum intrauterine pressure required for uterine distention. J Am Assoc Gynecol Laparosc. 1998;5:51‐53.
    1. Baker VL, Adamson GD. Threshold intrauterine perfusion pressures for intraperitoneal spill during hydrotubation and correlation with tubal adhesive disease. Fertil Steril. 1995;64:1066‐1069.
    1. Haggag HM, Hassan AMA. The impact of altering filling pressures in diagnostic outpatient hysteroscopy on the procedure completion rates and associated pain: a randomised double‐blind controlled trial. Aust NZ J Obstet Gynecol. 2016;56:97‐101.
    1. Sanchez Ibanez JM. Vascular factors in dysmenorrhea. Rev Esp Obstet Ginecol. 1950;9:104‐115.
    1. Vilos GA, Vilos EC, Pendley L. Endometrial ablation with a thermal balloon for the treatment of menorrhagia. J Am Assoc Gynecol Laparosc. 1996;3(3):383‐387.

Substances

LinkOut - more resources