Validation of combiomed hipermax-BF model A7101 automatic oscillometric upper-arm sphygmomanometer in general population: AAMI/ESH/ISO universal standard (ISO 81060-2:2018/Amd 1:2020)
- PMID: 39182003
- DOI: 10.1038/s41371-024-00948-9
Validation of combiomed hipermax-BF model A7101 automatic oscillometric upper-arm sphygmomanometer in general population: AAMI/ESH/ISO universal standard (ISO 81060-2:2018/Amd 1:2020)
Abstract
This study evaluates the accuracy of the Hipermax-BF model A7101 (Combiomed, Havana, Cuba) automatic oscillometric upper-arm sphygmomanometer for office and home use in general population as part of the HEARTS in the Americas initiative. The research was developed according to the Universal Standard AAMI/ESH/ISO ISO 81060-2:2018/Amd 1:2020. The subjects were recruited according to the requirements of age, gender, blood pressure values and upper-arm circumference. The same upper-arm sequential blood pressure measurement method was used. For measurements with the device under test, the 2-piece cuff from 22-44 cm limb circumference range was used. 92 subjects were recruited and 85 were analyzed. Mean age was 44.8 ± 14.7 years, mean upper-arm circumference was 32.3 ± 6.2, and 56.5% were female. For Validation Criterion 1, the mean value ± standard deviation of the differences in readings between the device under test and the reference device was 1.2 ± 4.9/0.8 ± 4.9 mmHg (systolic/diastolic). For both pressures, in criterion 1 the standard requires a mean value of the differences ≤ ± 5 mmHg and a standard deviation ≤ ± 8 mmHg. For Validation Criterion 2, the standard deviation of the mean blood pressure differences per subject was 4.2/4.2 mmHg (systolic/diastolic). According to Table 1 of criterion 2, for the mean values of 1.2/0.8 mmHg (systolic/diastolic), the maximum allowable standard deviation had to be < 6.84 for systolic and < 6.89 for diastolic pressure. The Combiomed Hipermax-BF A7101 automatic sphygmomanometer meets the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018/Amd 1:2020) in the general population.
© 2024. The Author(s), under exclusive licence to Springer Nature Limited.
References
-
- Campbell NRC, Schutte AE, Varghese CV, Ordunez P, Zhang X-H, Khan T, et al. São Paulo call to action for the prevention and control of high blood pressure: 2020. J Clin Hypertens. 2019;21:1744–52. - DOI
-
- Padwal R, Campbell NRC, Schutte AE, Olsen MH, Delles C, Etyang A, et al. Optimizing observer performance of clinic blood pressure measurement: a position statement from the Lancet Commission on Hypertension Group. J Hypertens. 2019;37:1737–45. https://doi.org/10.1097/HJH.0000000000002112 . - DOI - PubMed - PMC
-
- Padwal R, Berg A, Gelfer M, Tran K, Ringrose J, Ruzicka M, et al. Accuracy in measurement of blood pressure (AIM-BP) Collaborative. The Hypertension Canada blood pressure device recommendation listing: Empowering use of clinically validated devices in Canada. J Clin Hypertens (Greenwich). 2020;22:933–6. https://doi.org/10.1111/jch.13868 . - DOI - PubMed
-
- How to check that a blood pressure monitor has been properly tested for accuracy. https://www.menzies.utas.edu.au/documents/research/blood-pressure-clinic... .
-
- Picone DS, Deshpande RA, Schultz MG, Fonseca R, Campbell NRC, Delles C, et al. Nonvalidated home blood pressure devices dominate the online marketplace in Australia: major implications for cardiovascular risk management. Hypertension. 2020;75:1593–9. https://doi.org/10.1161/HYPERTENSIONAHA.120.14719 . - DOI - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous