Effect of Screw Length in Volar Plating for Intra-Articular Distal Radius Fractures: A Biomechanical Study

Affiliations
Free article
Comparative Study

Effect of Screw Length in Volar Plating for Intra-Articular Distal Radius Fractures: A Biomechanical Study

Thomas S Hong et al. J Hand Surg Am. 2024 Dec.
Free article

Abstract

Purpose: The purpose of this study was to compare the effect of varying screw lengths on load to failure and retention of the dorsal ulnar corner fragment after fixation of comminuted intra-articular distal radius fractures in a cadaveric model.

Methods: Twenty-four fresh frozen cadaveric forearms were subjected to a standardized distal radius osteotomy to mimic an intra-articular fracture pattern. Dual X-ray absorptiometry scans were performed to ensure minimal variability in bone density. All fractures were fixed with a volar locking plate and distal locking screws. Three different lengths of distal locking screws were used in each group of eight specimens to simulate the clinical decision of different distal screw lengths. The screw lengths tested were bicortical, 100% of the width of the bone but unicortical, and 75% of the width of the bone and unicortical. All specimens were preconditioned with cyclic axial loading and then axially loaded using matching acrylic resin molds to clinical failure and fragment displacement as detected by a motion analysis system. Retention or loss of the dorsal ulnar corner fragment during loading was recorded as a binary variable.

Results: Between the three groups, there were no statistically significant differences in precycling stiffness, postcycling stiffness, load at 2 mm displacement of the dorsal ulnar corner, or force at failure. The group with 75% length screws had a significantly higher loss of reduction of the dorsal ulnar corner (86%) compared with the other groups (0%).

Conclusions: Varying screw lengths did not affect the stiffness or overall loads to failure of axially loaded specimens. However, the 75% length screws did not reliably secure the dorsal ulnar corner fragments. Although this did not significantly affect the overall load to failure of the construct, displacement of this fragment may have implications for rotation of the forearm through the distal radioulnar joint.

Clinical relevance: Surgeons should consider the utilization of full-length unicortical locking screws to ensure adequate fixation of the dorsal ulnar corner.

Keywords: Biomechanics; distal radius fracture; dorsal ulnar corner; internal fixation.