. 2024 Dec;48(6):3889-3897.
doi: 10.1007/s11259-024-10555-1. Epub 2024 Sep 27.

Bioelectrical Impedance Analysis (BIA) detects body resistance increase in dogs undergoing blood donation

Affiliations

Bioelectrical Impedance Analysis (BIA) detects body resistance increase in dogs undergoing blood donation

Noemi Nisini et al. Vet Res Commun. 2024 Dec.

Abstract

Using raw parameters of bioelectrical impedance analysis (BIA) and derived Phase Angle (PhA) has emerged as an alternative to conventional predictive BIA for body fluid volumes and body composition estimation in human patients, offering an approach independent of predictive equations. Classic predictive equation-based BIA remains unsuitable for veterinary clinical practice due to the high morphological variability within the canine species. This study aimed to validate the feasibility of BIA in dogs and assess its utility in detecting changes in body fluid volumes, particularly in blood donation. Sixty healthy blood donor dogs and ten controls were longitudinally assessed using BIA raw parameters with a portable analyzer at a 50 kHz frequency. Measurements were performed before and after blood donation in donor dogs, whereas in controls, they were carried out at two time points, approximating the timing of a donation. Significant increases in raw bioelectrical parameters were observed after blood donation in the donor group, while differences were not statistically significant in controls. Hematocrit values did not significantly differ between the two time points in both groups. These findings suggest the feasibility of using BIA to detect changes in body fluid volumes in dogs undergoing blood donation, highlighting its potential application in veterinary medicine. Further research should explore the applicability of BIA in spontaneous pathological conditions involving fluid loss or retention and investigate the potential utility of PhA as a prognostic or monitoring tool in veterinary clinical practice. This study provides valuable insights into using BIA in canine medicine, paving the way for enhanced diagnostic and monitoring strategies in clinical practice.

Keywords: Bioelectrical impedance analysis; Blood donation; Dog; Phase angle; Total body water.

PubMed Disclaimer

Conflict of interest statement

The authors declare that the research was conducted without any commercial and financial relationship that could be construed as a potential conflict of interest.

Figures

Fig. 1
Fig. 1
Whole-body right tetrapolar electrode configuration for BIA analysis. Two emitting current-injecting electrodes (red and black) are placed dorsal to the elbow and dorsal to the patella. Two receiving voltage-sensing electrodes (blue and green) are applied 3.5 cm dorsally to their emitting electrodes. An imperceptible alternating sinusoidal electric current of 800 μA at an operating frequency of 50 kHz was used through the current-injecting distal pair
Fig. 2
Fig. 2
Biosmart ® EX.516(XX) calibration test

References

    1. Alves FD, Souza GC, Aliti GB et al (2015) Dynamic changes in bioelectrical impedance vector analysis and phase angle in acute decompensated heart failure. Nutrition 31:84–89. 10.1016/j.nut.2014.05.004 - PubMed
    1. Asogwa CO, Lai DTH (2017) A review on opportunities to assess hydration in wireless body area networks. Electronics 6:82. 10.3390/electronics6040082
    1. Barbosa-Silva MC, Barros AJD, Wang J et al (2005) Bioelectrical impedance analysis: population reference values for phase angle by age and sex. Am J Clin Nutr 82:49–52. 10.1093/ajcn/82.1.49 - PubMed
    1. Baumgartner RN, Chumlea WC, Roche AF (1988) Bioelectric impedance phase angle and body composition. Am J Clin Nutr 48:16–23. 10.1093/ajcn/48.1.16 - PubMed
    1. Bellido D, García-García C, Talluri A et al (2023) Future lines of research on phase angle: strengths and limitations. Rev Endocr Metab Disord 24:563–583. 10.1007/s11154-023-09803-7 - PMC - PubMed

LinkOut - more resources