Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2024 Oct 3;4(10):e0003760.
doi: 10.1371/journal.pgph.0003760. eCollection 2024.

Performance and usability of cardiometabolic point of care devices in Nepal: A prospective, quantitative, accuracy study

Affiliations

Performance and usability of cardiometabolic point of care devices in Nepal: A prospective, quantitative, accuracy study

Marina Giachino et al. PLOS Glob Public Health. .

Abstract

Non-communicable diseases (NCDs), such as cardiovascular disease and diabetes, represent a serious global health concern. There is an urgent need for prompt diagnosis and effective monitoring at point of care, especially in low- and middle-income countries. Here we present the results of a study assessing the quantitative accuracy of two devices that may fit the target product profile for a cardiometabolic point-of-care device. This prospective, quantitative, accuracy study (NCT05257564) was conducted between March to May 2022, investigating the performance of the JanaCare Aina Blood Monitoring System (JCAina) and the Tascom SimplexTAS 101 device (TAS101) compared with local standard laboratory methods in rural Nepal. Using fingerstick capillary blood, cardiometabolic parameters were analysed using both devices. The quantitative accuracy was compared against a local laboratory reference assay. System usability was also assessed. For JCAina, the mean absolute biases (Bland-Altman analysis) for glucose, HbA1c and total cholesterol tests were -3.87 mg/dL (95% CI: -7.52--0.22), 1.34% (95% CI: 1.21-1.47), and -9.52 mg/dL (95% CI: -11.9--7.2), respectively, corresponding to mean percentage biases of 2.0%, 18.5%, and -6.4%. These indicate clinically small (<10% biases) differences from laboratory results for glucose and cholesterol, and a moderate (10-20%) positive bias for HbA1c. For TAS101, the mean absolute biases for glucose, HbA1c, total cholesterol and creatinine tests were 18.7 mg/dL (95% CI: 15.8-21.5), -0.2% (95% CI: -0.26--0.14), 29.8 mg/dL (95% CI: 27.0-32.6), and -0.02 mg/dL (95% CI: -0.05-0.01), respectively, corresponding to mean percentage biases of 12.1%, -2.6%, 15.8%, and -4.5%. These indicate clinically small differences for HbA1c and creatinine, and moderate positive biases for glucose and cholesterol. Both systems exhibited usability challenges. The JCAina and TAS101 point-of-care cardiometabolic devices were shown to have promising accuracy in environmental conditions such as in Nepal, though improvements are still needed for some parameters and for ease of use. Trial registration: NCT05257564 (ClinicalTrials.gov).

PubMed Disclaimer

Conflict of interest statement

Marina Giachino, Zoltan Pataky, Olivia Heller, Bruno Lab, Jorge César Correia, Sanjib Kumar Sharma, Vijay Kumar Khanal, Sagar Poudel, Mamit Rai declared that no competing interests exist. I have read the journal’s policy and the following authors: Beatrice Vetter, Berra Erkosar have the following competing interests: they are both employees of FIND. As for the author: Sigiriya Aebischer Perone, she has the following competing interests: is an employee of ICRC, the International Committee of the Red Cross.

Figures

Fig 1
Fig 1. Patient and testing disposition.
Hb, haemoglobin; HbA1c, glycated haemoglobin.
Fig 2
Fig 2. Passing Bablok analysis of JCAina tests (glucose, HbA1c, total cholesterol).
HbA1c, glycosylated haemoglobin; JCAina, JanaCare Aina Blood Monitoring System.
Fig 3
Fig 3. Bland-Altman analysis of JCAina tests (glucose, HbA1c, total cholesterol).
HbA1c, glycosylated haemoglobin; JCAina, JanaCare Aina Blood Monitoring System.
Fig 4
Fig 4. Passing Bablok analysis of TAS101 tests (glucose, HbA1c, total cholesterol, creatinine).
HbA1c, glycosylated haemoglobin; TAS101, Tascom SimplexTAS 101 device.
Fig 5
Fig 5. Bland-Altman analysis of TAS101 tests (glucose, HbA1c, total cholesterol, creatinine).
HbA1c, glycosylated haemoglobin; TAS101, Tascom SimplexTAS 101 device.

References

    1. World Health Organization (WHO). Noncommunicable diseases. Available from https://www.who.int/en/news-room/fact-sheets/detail/noncommunicable-dise... (accessed 01 Feb 2023). 2022.
    1. FIND. Landscape of point-of-care devices for testing of cardiometabolic diseases. Available from https://www.finddx.org/wp-content/uploads/2023/02/20200228_rep_landscape... (accessed 26 March 2023). 2020.
    1. World Health Organization (WHO). Package of essential noncommunicable (PEN) disease interventions for primary health care in low resource settings. Available from https://apps.who.int/iris/rest/bitstreams/1301957/retrieve (accessed 02 February 2023). 2020.
    1. World Health Organization (WHO). The selection and use of essential in vitro diagnostics. Available from https://apps.who.int/iris/rest/bitstreams/1329518/retrieve (accessed 02 February 2023). 2021.
    1. Albitres-Flores L, Perez-Leon S, Bernabe-Ortiz A, Tenorio-Mucha J, Cardenas MK, Vetter B, et al. Co-creation process of an intervention to implement a multiparameter point-of-care testing device in a primary healthcare setting for non-communicable diseases in Peru. BMC Health Services Research. 2024;24(1):401. doi: 10.1186/s12913-024-10809-3 - DOI - PMC - PubMed

Associated data

LinkOut - more resources