Effectiveness of the Intrauterine Balloon Tamponade Compared With an Intrauterine, Vacuum-Induced, Hemorrhage-Control Device for Postpartum Hemorrhage
- PMID: 39447181
- DOI: 10.1097/AOG.0000000000005770
Effectiveness of the Intrauterine Balloon Tamponade Compared With an Intrauterine, Vacuum-Induced, Hemorrhage-Control Device for Postpartum Hemorrhage
Abstract
Objective: To compare intrauterine balloon tamponade and vacuum-induced hemorrhage-control devices in a multicenter hospital system using a standardized three-stage postpartum hemorrhage response algorithm.
Methods: As part of a quality improvement project, data were prospectively collected through ongoing detailed hemorrhage case audits from 65 facilities from August 2022 to February 2024. Postpartum hemorrhage-control devices are recommended for stage 2 or 3 of the postpartum hemorrhage algorithm. The primary outcomes were 1) quantitative blood loss after device placement, 2) rate of packed red blood cell (RBC) transfusion, 3) use of 3 units of packed RBCs or more, and 4) device failure.
Results: During the 17-month observation period, there were 123,292 deliveries, and postpartum hemorrhage occurred in 5,931 (4.8%). Postpartum hemorrhage-control devices were used in 11.2% of cases (666 total, intrauterine balloon tamponade n=300, and vacuum-induced hemorrhage control n=366). For intrauterine balloon tamponade and vacuum-induced hemorrhage-control devices, quantitative blood loss (median and interquartile range) after device placement was similar at 194 (67-440) mL and 240 (113-528) mL ( P =.40), respectively, as was the rate of any packed RBC transfusion (59.7% vs 50.0%, P =.08), transfusion of 3 packed RBC units or more (27.0% vs 24.9%, P =.53), and device failure (7.7% vs 8.5%, P =.70). Placement of either device at a quantitative blood loss between 1,000 and 1,499 mL compared with 1,500 mL or more resulted in significantly lower rates of packed RBC transfusion (39.1% vs 70.3%, P <.001), transfusion of 3 or more packed RBC units (13.7 vs 38.1%, P <.001), and device failure (3.4% vs 12.9%, P <.001).
Conclusion: Transfusion and blood loss after device placement were similar with both hemorrhage-control devices. Earlier hemorrhage-control device placement reduced device failure and the need for transfusion.
Copyright © 2024 by the American College of Obstetricians and Gynecologists. Published by Wolters Kluwer Health, Inc. All rights reserved.
Conflict of interest statement
Financial Disclosure The authors did not report any potential conflicts of interest.
References
-
- Postpartum hemorrhage. Practice Bulletin No. 183. American College of Obstetricians and Gynecologists. Obstet Gynecol 2017;130:e168–86. doi: 10.1097/AOG.0000000000002351 - DOI
-
- Callaghan WM, Kuklina EV, Berg CJ. Trends in postpartum hemorrhage: United States, 1994-2006. Am J Obstet Gynecol 2010;202:353.e1–6. doi: 10.1016/j.ajog.2010.01.011 - DOI
-
- Khan KS, Wojdyla D, Say L, Gulmezoglu AM, Van Look PF. WHO analysis of causes of maternal death: a systematic review. Lancet 2006;367:1066–74. doi: 10.1016/S0140-6736(06)68397-9 - DOI
-
- Jackson TL, Tuuli MG. Intrauterine postpartum hemorrhage-control devices. Obstet Gynecol 2023;142:1000–5. doi: 10.1097/AOG.0000000000005403 - DOI
-
- Centers for Disease Control and Prevention. Pregnancy Mortality Surveillance System. Accessed April 1, 2024. https://www.cdc.gov/maternal-mortality/php/pregnancy-mortality-surveilla...
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials