Diagnostic Approach to Suspected Lead Failure
- PMID: 39461827
- DOI: 10.1016/j.ccep.2024.06.003
Diagnostic Approach to Suspected Lead Failure
Abstract
Transvenous lead failure associated with cardiac pacing and defibrillation remains an important clinical problem, with an estimated incidence between 1 to 2%. Oversensing of non-physiological signals usually precede lead impedance changes and may result in clinical compliations such as pacing inhibition and inappropriate shocks. Device based algorithms that identify non-physiological signals can be used in conjunction with remote monitoring to facilitate early diagnosis and management of lead failure and avoid serious adverse outcomes. This review highlights mechanisms of lead failure and proposes a diagnostic approach to suspected lead failure.
Keywords: ICD; Lead; Oversensing; Pacing; Remote monitoring.
Copyright © 2024 Elsevier Inc. All rights reserved.
Conflict of interest statement
Disclosure None declared.