A multicenter study comparing the bacterial reduction on flexible endoscopes without a working channel between UV-C light disinfection versus standard endoscope Washer Disinfection: a randomized controlled trial
- PMID: 39462387
- PMCID: PMC11515285
- DOI: 10.1186/s13756-024-01486-2
A multicenter study comparing the bacterial reduction on flexible endoscopes without a working channel between UV-C light disinfection versus standard endoscope Washer Disinfection: a randomized controlled trial
Abstract
Background: To prevent cross-contamination between patients, adequate reprocessing is necessary when using flexible endoscopes (FEs) without a working channel. The current reprocessing process using an Endoscope Washer Disinfector (EWD) is time-consuming. Ultraviolet light group C (UV-C) exposition is an alternative and fast disinfection method and has previously been shown to adequately reduce Colony Forming Units (CFUs) on FEs without a working channel. The objective of this study was to examine whether UV-C light is as effective in reducing CFUs on contaminated FEs without a working channel compared to the EWD.
Methods: FEs without a working channel were collected in three different Otorhinolaryngology Departments in the Netherlands. After pharyngolaryngoscopy, a manual pre-cleaning with tap water was performed and a culture was collected by rolling the distal 8-10 cm of the FE over an agar plate. Next, the FE was randomly assigned to be disinfected with UV-C light (D60) or the EWD (gold standard). After disinfection, another culture was taken. The primary outcome was microbiological contamination, defined by Colony Forming Units (CFU).
Results: 600 FEs without a working channel were randomized. After clinical use and manual pre-cleaning, 239/300 (79.7%) FEs in the UV-C group and 262/300 (87.3%) FEs in the EWD group were contaminated (i.e., > 0 CFU). FEs without culture confirmed contamination were excluded from further analysis. After UV-C light disinfection, 195/239 (81.6%) FEs showed 0 CFUs, compared to 187/262 (71.4%) FEs disinfected with the EWD (p < 0.01). A multivariate logistics regression analysis showed an increased odds of 0 CFUs when using UV-C light (OR 1.83, 95% CI 1.19-2.79; p < 0.01), conditional on participating hospitals and types of FE.
Conclusions: UV-C light disinfection of FEs without a working channel appears more effective in reducing CFUs compared to the EWD and might be a good alternative disinfection method.
Trial registration: Not applicable.
Keywords: Colony forming units; Disinfection; Endoscope Washer Disinfector; Flexible endoscope without a working channel; UV-C; Ultraviolet light C.
© 2024. The Author(s).
Conflict of interest statement
The authors declare no competing interests.
References
-
- Seoane-Vazquez E, Rodriguez-Monguio R. Endoscopy-related infection: relic of the past? Curr Opin Infect Dis. 2008;21(4):362–6. - PubMed
-
- Spaulding EH. Disinfection, sterilization and preservation. In: Chemical disinfect of medical and surgical materials; 1968. pp. 517–31.
-
- Veerabadran S, Parkinson IM. Cleaning, disinfection and sterilization of equipment. Anaesth Intensive care Med. 2010;11(11):451–4.
-
- Muscarella LF. High-level disinfection or sterilization of endoscopes? Infect Control Hosp Epidemiol. 1996;17(3):183–7. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous