Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2025 Apr:198:141-147.
doi: 10.1016/j.urology.2024.11.043. Epub 2024 Nov 26.

Contrast-enhanced Ultrasound Imaging Following Transperineal Laser Ablation for Lower Urinary Tract Symptoms

Affiliations
Free article

Contrast-enhanced Ultrasound Imaging Following Transperineal Laser Ablation for Lower Urinary Tract Symptoms

R A A van Kollenburg et al. Urology. 2025 Apr.
Free article

Abstract

Objective: To describe the shape and volume of ablations created by transperineal laser ablation (TPLA) using multiple fiber configurations. Furthermore, to measure the change in the ablation zone and prostate volume over time, and to assess inter-patient ablation volume variability.

Methods: Data from a prospective, single-center, interventional pilot study including 20 patients is used. All subjects underwent TPLA using the EchoLaser system, using 2 to 4 fibers, depending on prostate size and shape. Contrast-enhanced ultrasound (CEUS) was performed post-treatment and at 1 and 12 months. The prostate and ablation zone volumes were calculated on segmented CEUS imaging.

Results: The ablation zones were clearly identified on CEUS as non-perfused areas. Depending on fiber configuration, their shape varied from an ellipsoid to a clover profile. Ablation volumes varied from 0.9 (0.6-2.2)cm3 using a single fiber and 1800 J to 8.7 (3.9-19.0)cm3 (median, range) using 2 fibers and 7200 J energy per lobe at 1 month. At 12 months, the majority of the ablation zones showed a volume reduction. Median prostate volume decreased from 78 (37-145)cm3 at baseline to 46 (27-124)cm3 at 12 months (P=.0002). There was a relation between prostate volume reduction and Qmax (slope=0.18) and IPSS (slope=-0.18) improvement.

Conclusion: This study described ablation zone shape and measured the ablation volume following TPLA by various fiber configurations using CEUS, and compared these to functional outcomes. Prostate volume reduced significantly during follow-up. Segmentation showed substantial inter-patient ablation volume variation, which limits treatment predictability and thus accuracy.

Trial registration: ClinicalTrials.gov NCT03653117.

PubMed Disclaimer

Conflict of interest statement

Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Comment in

References

Publication types

Associated data