Degradation and Fatigue Behavior of 3D-Printed Bioresorbable Tracheal Splints
- PMID: 39607369
- PMCID: PMC11619926
- DOI: 10.1002/jbm.b.35501
Degradation and Fatigue Behavior of 3D-Printed Bioresorbable Tracheal Splints
Abstract
Severe infantile tracheobronchomalacia (TBM) is often treated with invasive surgery and fixed-size implants to support the trachea during respiration. A novel 3D-printed extra-luminal splint has been developed as a flexible and bioresorbable alternative. Therefore, the goal of the present study was to use an in vitro breathing simulator model to comprehensively evaluate the structural stiffness and failure modes of two sizes of a novel bioresorbable 3D-printed splint design under a range of physiological degradation conditions. Two thicknesses, 2 mm and 3 mm, of a novel 3D-printed bioresorbable splint were evaluated under two different degradation conditions, phosphate-buffered saline (PBS) and sodium hydroxide (NaOH). The splints were subjected to simulated breathing loading, involving a cyclic opening and closing of the splint by 2 mm, for a targeted duration of 7.5 to 30 million cycles. A separate new set of splints were statically soaked in their respective degradation condition for a comparative analysis of the effects of cyclic loading by the degradation medium. After successfully simulated breathing or static soaking, non-destructive tensile and compressive strengths were evaluated, and overall stiffness was calculated from destructive tensile testing. The present study indicates that the splints were more significantly degraded under simulated breathing conditions than under soaking. Cyclic simulated breathing specimens failed far earlier than the intended duration of loading. Over time, both 2 mm and 3 mm splints became increasingly more flexible when subjected to the static degradation conditions. Interestingly, there was little difference in the compressive and tensile strengths of the 2 mm and 3 mm thickness splints. The bioresorbable nature of PCL offers a valuable advantage as it eliminates the need for splint removal surgery and increases device flexibility over time with degradation. This increased flexibility is crucial because it allows for uninhibited growth and development of the infant's trachea over the intended use period of 2 years. The results of this study confirm that the splints were able to withstand tensile forces to prevent tracheal collapse. This study further supports the successful use of 3D-printed splints in the treatment of infantile TBM.
Keywords: 3D‐printing; Tracheobronchomalacia; degradation; fatigue; splint.
© 2024 The Author(s). Journal of Biomedical Materials Research Part B: Applied Biomaterials published by Wiley Periodicals LLC.
Similar articles
-
Mg-Zn-Ca Alloy (ZX00) Screws Are Resorbed at a Mean of 2.5 Years After Medial Malleolar Fracture Fixation: Follow-up of a First-in-humans Application and Insights From a Sheep Model.Clin Orthop Relat Res. 2024 Jan 1;482(1):184-197. doi: 10.1097/CORR.0000000000002799. Epub 2023 Aug 21. Clin Orthop Relat Res. 2024. PMID: 37603369 Free PMC article.
-
Design, fabrication and comprehensive testing of biodegradable 3D printable hybrid polymer airway splints.J Mater Chem B. 2025 Aug 20;13(33):10249-10266. doi: 10.1039/d5tb00891c. J Mater Chem B. 2025. PMID: 40734618
-
The Black Book of Psychotropic Dosing and Monitoring.Psychopharmacol Bull. 2024 Jul 8;54(3):8-59. Psychopharmacol Bull. 2024. PMID: 38993656 Free PMC article. Review.
-
3D-Printed anterior repositioning splint versus stabilization splint for patients with anterior disc displacement with reduction: A randomized cross-over clinical trial.J Indian Prosthodont Soc. 2025 Apr 1;25(2):163-168. doi: 10.4103/jips.jips_8_25. Epub 2025 Apr 11. J Indian Prosthodont Soc. 2025. PMID: 40213888 Free PMC article. Clinical Trial.
-
Guided tissue regeneration for periodontal infra-bony defects.Cochrane Database Syst Rev. 2006 Apr 19;(2):CD001724. doi: 10.1002/14651858.CD001724.pub2. Cochrane Database Syst Rev. 2006. Update in: Cochrane Database Syst Rev. 2019 May 29;5:CD001724. doi: 10.1002/14651858.CD001724.pub3. PMID: 16625546 Updated.
References
-
- Murgu SD, Colt HG. Tracheobronchomalacia and excessive dynamic airway collapse. Respirology 2006; 11:388–406. - PubMed
-
- Boogaard R, Huijsmans SH, Pijnenburg MWH, Tiddens HAWM, de Jongste JC, Merkus PJFM. Tracheomalacia and Bronchomalacia in Children: Incidence and Patient Characteristics. Chest 2005; 128:3391–3397. - PubMed
-
- Boiselle PM, Feller-Kopman D, Ashiku S, Weeks D, Ernst A. Tracheobronchomalacia: evolving role of dynamic multislice helical CT. Radiol Clin North Am 2003; 41:627–636. - PubMed
-
- Jacobs IN, Wetmore RF, Tom LW, Handler SD, Potsic WP. Tracheobronchomalacia in children. Arch Otolaryngol Head Neck Surg 1994; 120:154–158. - PubMed
-
- Lee EY, Boiselle PM. Tracheobronchomalacia in infants and children: multidetector CT evaluation. Radiology 2009; 252:7–22. - PubMed