Performance Evaluation of a Coagulation Laboratory Using Sigma Metrics and Quality Goal Index
- PMID: 39662009
- DOI: 10.7754/Clin.Lab.2024.240619
Performance Evaluation of a Coagulation Laboratory Using Sigma Metrics and Quality Goal Index
Abstract
Background: Sigma methodology is a statistical calculation and quality management tool that provides information about process performance. If clinical laboratories start using sigma metrics to monitor their performance, they can more easily identify gaps in their performance, thereby improve their performance and patient safety. This study aimed to calculate sigma metrics and quality target index values by using internal quality control data from coagulation tests, thus evaluating the analytical performance.
Methods: Sigma levels were calculated using the formula: [total allowable error (TEa) - bias]/coefficient of variation (CV). Sigma values ≥ 6, between 3 and 6, and < 3 were classified as "world-class", "good" or "unacceptable", respectively. A biological variation database (BVD) was used for TEa. The quality goal index (QGI) is the reason behind a low sigma value, that is, lower precision, lower accuracy, or both due to the combination. QGI was calcu-lated using the formula QGI = bias/1.5 x % CV. With a QGI value of < 0.8, the measurement indicates that the accuracy of the procedure needs to be improved; QGI values > 1.2 indicate accuracy needs to be improved and values 0.8 ≤ QGI ≤ 1.2 indicate both precision and accuracy need to be improved.
Results: Sigma and QGI of three-monthly two-level internal quality control values were calculated by using the laboratory automation system. In the prothrombin time (PT) and activated partial thromboplastin time (APTT) tests of the coagulation parameters studied, sigma values were found to be < 3 in both levels. When the QGI value was calculated, it was PT 0.45 and APTT 0.90 for level 1 and PT 0.16 and APTT 0.6 for level 2, respectively.
Conclusions: It was decided that sigma values of coagulation parameters at "low quality" levels and improvement studies should be carried out for coagulation parameters in our laboratory. By evaluating sigma levels, it is possible to identify tests with a high probability of failure, and these tests should undergo strict quality control inspections. In clinical biochemistry laboratories, appropriate quality control planning should be performed for each test by using the Six Sigma methodology and by calculating the quality target index.
Similar articles
-
Beyond the basics: Sigma scores in laboratory medicine with variable total allowable errors (TEa).Clin Chim Acta. 2025 Jan 15;565:119971. doi: 10.1016/j.cca.2024.119971. Epub 2024 Sep 24. Clin Chim Acta. 2025. PMID: 39326693
-
The Application of Six Sigma Techniques in the Evaluation of Enzyme Measurement Procedures in China.Clin Lab. 2015;61(5-6):461-5. doi: 10.7754/clin.lab.2014.140915. Clin Lab. 2015. PMID: 26118176
-
Achieving High Standards in Clinical Biochemistry: Integrating Six Sigma, Quality Goal Index (QGI), and Operating Specifications (OPSpecs) for Targeted Quality Enhancement.Cureus. 2024 Nov 28;16(11):e74693. doi: 10.7759/cureus.74693. eCollection 2024 Nov. Cureus. 2024. PMID: 39735008 Free PMC article.
-
Sigma metrics in laboratory medicine revisited: We are on the right road with the wrong map.Biochem Med (Zagreb). 2018 Jun 15;28(2):020503. doi: 10.11613/BM.2018.020503. Biochem Med (Zagreb). 2018. PMID: 30022880 Free PMC article. Review.
-
Evidence-based assessment of the application of Six Sigma to infectious disease serology quality control.Clin Chem Lab Med. 2025 Feb 19;63(6):1228-1236. doi: 10.1515/cclm-2024-1455. Print 2025 May 26. Clin Chem Lab Med. 2025. PMID: 39964186 Review.
MeSH terms
LinkOut - more resources
Full Text Sources