Generative Adversarial Network Based Contrast Enhancement: Synthetic Contrast Brain Magnetic Resonance Imaging
- PMID: 39694785
- DOI: 10.1016/j.acra.2024.11.021
Generative Adversarial Network Based Contrast Enhancement: Synthetic Contrast Brain Magnetic Resonance Imaging
Abstract
Rationale and objectives: Magnetic resonance imaging (MRI) is a vital tool for diagnosing neurological disorders, frequently utilising gadolinium-based contrast agents (GBCAs) to enhance resolution and specificity. However, GBCAs present certain risks, including side effects, increased costs, and repeated exposure. This study proposes an innovative approach using generative adversarial networks (GANs) for virtual contrast enhancement in brain MRI, with the aim of reducing or eliminating GBCAs, minimising associated risks, and enhancing imaging efficiency while preserving diagnostic quality.
Material and methods: In this study, 10,235 images were acquired in a 3.0 Tesla MRI scanner from 81 participants (54 females, 27 males; mean age 35 years, range 19-68 years). T1-weighted and contrast-enhanced images were obtained following the administration of a standard dose of a GBCA. In order to generate "synthetic" images for contrast-enhanced T1-weighted, a CycleGAN model, a sub-model of the GAN structure, was trained to process pre- and post-contrast images. The dataset was divided into three subsets: 80% for training, 10% for validation, and 10% for testing. TensorBoard was employed to prevent image deterioration throughout the training phase, and the image processing and training procedures were optimised. The radiologists were presented with a non-contrast input image and asked to choose between a real contrast-enhanced image and synthetic MR images generated by CycleGAN corresponding to this non-contrast MR image (Turing test).
Results: The performance of the CycleGAN model was evaluated using a combination of quantitative and qualitative analyses. For the entire dataset, in the test set, the mean square error (MSE) was 0.0038, while the structural similarity index (SSIM) was 0.58. Among the submodels, the most successful model achieved an MSE of 0.0053, while the SSIM was 0.8. The qualitative evaluation was validated through a visual Turing test conducted by four radiologists with varying levels of clinical experience.
Conclusion: The findings of this study support the efficacy of the CycleGAN model in generating synthetic contrast-enhanced T1-weighted brain MR images. Both quantitative and qualitative evaluations demonstrated excellent performance, confirming the model's ability to produce realistic synthetic images. This method shows promise in potentially eliminating the need for intravenous contrast agents, thereby minimising the associated risks of their use.
Keywords: Artificial Intelligence; Augmented Contrast; Gadolinium-Based Contrast Agents; Generative Adversarial Networks; MRI; Neuro Imaging; Synthetic Imaging; Virtual Contrast.
Copyright © 2025 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.
Conflict of interest statement
Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical