Construction and performance of a variable-frequency phase-modulation fluorometer
- PMID: 3971026
- DOI: 10.1016/0301-4622(85)85007-9
Construction and performance of a variable-frequency phase-modulation fluorometer
Abstract
We describe the construction and performance of a variable-frequency phase-modulation fluorometer. This instrument, which provides modulation frequencies from 1 to 200 MHz, was constructed using commercially available components. To facilitate the introduction of these instruments into other laboratories we describe in detail the chosen components and the principles of operation. The present light source is a continuous-wave helium-cadmium laser, which provides convenient excitation wavelengths of 325 and 442 nm. Modulation of the incident light is provided by one of several electro-optic modulators. The extent of modulation ranges from 1.0 to 0.2 as the frequency increases from 1 to 200 MHz. Phase angles and demodulation factors are measured using the cross-correlation method. The closely spaced frequencies are provided by two direct frequency synthesizers. The phase and modulation measurements are accurate to 0.2 degrees and 0.002, respectively, from 1 to 200 MHz. This accuracy allows considerable resolution of complex decay laws. The usefulness of frequency-domain fluorometry for the resolution of multiexponential decays is illustrated by the analysis of several difficult mixtures. As examples, we resolved a two-component mixture of anthracene (4.1 ns) and 9,10-diphenylanthracene (6.3 ns), and confirmed that the intensity decay of NADH in aqueous buffer is at least a double exponential (0.2 and 0.86 ns). We also resolved an especially difficult mixture of anthracene (4.1 ns) and 9-methylanthracene (4.5 ns), and a three-component mixture with decay times of 1.3, 4.1 and 7.7 ns. Frequency-domain fluorometers appear to be particularly useful for determination of complex decays of fluorescence anisotropy. This capability is illustrated by the determination of rotational correlation times as short as 47 ps for p-bis[2-(5-phenyloxazolyl)]benzene (POPOP) in hexane at 40 degrees C, and by the resolution of the two correlation times of anisotropic rotators such as perylene and 9-aminoacridine. Resolution of two anisotropy decay times for 9-aminoacridine is a difficult test because these correlation times differ by less than 2-fold. The resolution of multiexponential decays of intensity and anisotropy possible with this instrument is at least equivalent to that obtained using state-of-the-art time-resolved instruments based on mode-locked laser sources. The ease and rapidity of frequency-domain measurements, the relative simplicity of the equipment, the accuracy of the measurements and the lack of significant systematic errors indicate that frequency-domain fluorometry will be widely useful in chemical and biochemical research.
Similar articles
-
A new method for resolution of two- and three-component mixtures of fluorophores by phase-sensitive detection of fluorescence.Anal Biochem. 1985 Aug 1;148(2):349-56. doi: 10.1016/0003-2697(85)90239-8. Anal Biochem. 1985. PMID: 4061815
-
Resolution of mixtures of fluorophores using variable-frequency phase and modulation data.Biophys J. 1984 Oct;46(4):479-86. doi: 10.1016/S0006-3495(84)84044-8. Biophys J. 1984. PMID: 6498265 Free PMC article.
-
Time-resolved fluorescence anisotropies of diphenylhexatriene and perylene in solvents and lipid bilayers obtained from multifrequency phase-modulation fluorometry.Biochemistry. 1985 Jan 15;24(2):376-83. doi: 10.1021/bi00323a021. Biochemistry. 1985. PMID: 3978080 Free PMC article.
-
Frequency domain fluorometry with pulsed light-emitting diodes.Ann N Y Acad Sci. 2008;1130:56-61. doi: 10.1196/annals.1430.026. Ann N Y Acad Sci. 2008. PMID: 18596332 Review.
-
Gigahertz frequency-domain fluorometry: resolution of complex decays, picosecond processes and future developments.J Photochem Photobiol B. 1988 Nov;2(3):295-311. doi: 10.1016/1011-1344(88)85050-4. J Photochem Photobiol B. 1988. PMID: 3148695 Review.
Cited by
-
Distance distributions and anisotropy decays of troponin C and its complex with troponin I.Biochemistry. 1991 May 28;30(21):5238-47. doi: 10.1021/bi00235a018. Biochemistry. 1991. PMID: 2036391 Free PMC article.
-
Conformation of the C1 phorbol-ester-binding domain participates in the activating conformational change of protein kinase C.Biochem J. 1999 Dec 1;344 Pt 2(Pt 2):451-60. Biochem J. 1999. PMID: 10567228 Free PMC article.
-
DNA dynamics: a fluorescence resonance energy transfer study using a long-lifetime metal-ligand complex.Arch Pharm Res. 2002 Apr;25(2):143-50. doi: 10.1007/BF02976554. Arch Pharm Res. 2002. PMID: 12009026 Free PMC article.
-
Long-lifetime metal-ligand pH probe.Anal Biochem. 1997 May 1;247(2):216-22. doi: 10.1006/abio.1997.2057. Anal Biochem. 1997. PMID: 9177680 Free PMC article.
-
Fluorescence detection of the anticancer drug topotecan in plasma and whole blood by two-photon excitation.Anal Biochem. 1996 Nov 15;242(2):266-70. doi: 10.1006/abio.1996.0463. Anal Biochem. 1996. PMID: 8937572 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials
Miscellaneous