Directional vector-based quick evaluation method for protective plate effects in X-ray fluoroscopy (DQPEX)
- PMID: 39733369
- DOI: 10.1007/s12194-024-00873-z
Directional vector-based quick evaluation method for protective plate effects in X-ray fluoroscopy (DQPEX)
Abstract
One radiation protection measure for medical personnel in X-ray fluoroscopy is using radiation protective plates. A real-time interactive tool visualizing radiation-dose distribution varied with the protective plate position will help greatly to train medical personnel to protect themselves from unnecessary radiation exposure. Monte Carlo simulation can calculate the individual interactions between radiations and objects in the X-ray room, and reproduce the complex dose distribution inside the room. However, Monte Carlo simulation is computationally time-consuming and not suited for real-time feedback. Therefore, we developed a new method to calculate the dose distribution with the presence of protective plates instantly using pre-computed directional vectors, named Directional vector-based Quick evaluation method for Protective plates Effects in X-ray fluoroscopy (DQPEX). DQPEX uses a database of dose distributions and directional vectors precomputed by Monte Carlo code, Particle and Heavy Ion Transport code System (PHITS). Assuming the dose at each position was all contributed from radiations in the direction indicated by the directional vector, the dose reduction by the protective plates at the position was determined whether the backtrace line of the directional vector has a intersect with the protective plate or not. With DQPEX, the whole dose distribution in X-ray room with the presence of a protective plate can be computed about 13 s, which is approximately 1/6000 of the full PHITS simulation. Sufficient accuracy of DQPEX to visualize the effect of a protective plate was confirmed by comparing the obtained dose distribution with those obtained by the full PHITS simulation and measurements.
Keywords: Directional vector; Occupational exposure; Protective plate; Scattered radiation; Visualization; X-ray fluoroscopy.
© 2024. The Author(s), under exclusive licence to Japanese Society of Radiological Technology and Japan Society of Medical Physics.
Conflict of interest statement
Declarations. Conflict of interest: The authors declare no conflicts of interest. Ethics approval: This article does not include any studies performed on human participants or animals.
Similar articles
-
Directional vector visualization of scattered rays in mobile c-arm fluoroscopy.Radiol Phys Technol. 2024 Mar;17(1):288-296. doi: 10.1007/s12194-024-00779-w. Epub 2024 Feb 5. Radiol Phys Technol. 2024. PMID: 38316688
-
[Development of Educational Materials for Visualizing the Directional Vectors of Scattered Radiation during X-ray Fluoroscopy Using Virtual Reality Technology].Nihon Hoshasen Gijutsu Gakkai Zasshi. 2025;81(9). doi: 10.6009/jjrt.25-1573. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2025. PMID: 40571627 Japanese.
-
Visualization of spatial dose distribution for effective radiation protection education in interventional radiology: obtaining high-accuracy spatial doses.Phys Eng Sci Med. 2024 Dec;47(4):1665-1676. doi: 10.1007/s13246-024-01479-w. Epub 2024 Sep 9. Phys Eng Sci Med. 2024. PMID: 39249663 Free PMC article.
-
Radiation Protection for the Fluoroscopy Operator and Staff.AJR Am J Roentgenol. 2016 Oct;207(4):745-754. doi: 10.2214/AJR.16.16556. Epub 2016 Jul 19. AJR Am J Roentgenol. 2016. PMID: 27440524 Review.
-
Track-structure modes in particle and heavy ion transport code system (PHITS): application to radiobiological research.Int J Radiat Biol. 2022;98(2):148-157. doi: 10.1080/09553002.2022.2013572. Epub 2021 Dec 20. Int J Radiat Biol. 2022. PMID: 34930091 Review.
References
-
- Miller DL. Make radiation protection a habit. Tech Vasc Interv Radiol. 2018;21(1):37–42. https://doi.org/10.1053/j.tvir.2017.12.008 . - DOI - PubMed
-
- ICRP. Radiological protection in fluoroscopically guided procedures outside the imaging department. Ann ICRP. 2010. https://doi.org/10.1016/j.icrp.2012.03.001 . - DOI
-
- Abdelrahman M, Lombardo P, Camp A, et al. A parametric study of occupational radiation dose in interventional radiology by Monte-Carlo simulations. Phys Med. 2020;78:58–70. https://doi.org/10.1016/j.ejmp.2020.08.016 . - DOI - PubMed
-
- Miller DL. Overview of contemporary interventional fluoroscopy procedures. Health Phys. 2008;95(5):638–44. https://doi.org/10.1097/01.HP.0000326341.86359.0b . - DOI - PubMed
-
- Vano E, Gonzalez L, Fernández JM, Haskal ZJ. Eye lens exposure to radiation in interventional suites: caution is warranted. Radiology. 2008;248(3):945–53. https://doi.org/10.1148/radiol.2482071800 . - DOI - PubMed
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources