Artifact Reduction in Interventional Devices Using Virtual Monoenergetic Images and Iterative Metal Artifact Reduction on Photon-Counting Detector CT
- PMID: 39745893
- DOI: 10.1097/RLI.0000000000001149
Artifact Reduction in Interventional Devices Using Virtual Monoenergetic Images and Iterative Metal Artifact Reduction on Photon-Counting Detector CT
Abstract
Objectives: The aim of this study was to assess the impact of an iterative metal artifact reduction (iMAR) algorithm combined with virtual monoenergetic images (VMIs) for artifact reduction in photon-counting detector computed tomography (PCDCT) during interventions.
Materials and methods: Using an abdominal phantom, we conducted evaluations on the efficacy of iMAR and VMIs for mitigating image artifacts during interventions on a PCDCT. Four different puncture devices were employed under 2 scan modes (QuantumSn at 100 kV, Quantumplus at 140 kV) to simulate various clinical scenarios. Image reconstructions were initially performed without iMAR and subsequently with iMAR settings. The latter was tested with 7 different metal presets for each case. Furthermore, iMAR-reconstructed images were paired with VMIs at energy levels of 70 keV, 110 keV, 150 keV, and 190 keV. Qualitative assessments were conducted to evaluate image quality, artifact expression, and the emergence of new artifacts using a Likert scale. Image quality was rated on a scale of 1 (nondiagnostic) to 5 (excellent), whereas artifact severity was rated from 0 (none) to 5 (massive). Preferences for specific iMAR presets were documented. Quantitative analysis involved calculating Hounsfield unit (HU) differences between artifact-rich and artifact-free tissues.
Results: Overall, 96 different scanning series were evaluated. The optimal combination for artifact reduction was found to be iMAR neurocoils with VMIs at 150 keV and 190 keV, showcasing the most substantial reduction in artifacts with a median rating of 1 (standard: 4). VMIs at higher keV levels, such as 190 keV, resulted in reduced image quality, as indicated by a median rating of 3 (compared with 70 keV with a median of 5). Newly emerged artifact expression related to reconstructions varied among intervention devices, with iMAR thoracic coils exhibiting the least extent of artifacts (median: 2) and iMAR neurocoils displaying the most pronounced artifacts (median: 4). Qualitative analysis favored the combination of iMAR neurocoils with VMIs at 70 keV, showcasing the best results. Conversely, quantitative analysis revealed that the combination of iMAR neurocoils with VMIs at 190 keV yielded the best results, with an average artifact expression of 20.06 HU (standard: 167.98 HU; P < 0.0001).
Conclusions: The study underscores a substantial reduction in artifacts associated with intervention devices during PCDCT scans through the synergistic application of VMI and iMAR techniques. Specifically, the combination of VMIs at 70 keV with iMAR neurocoils was preferred, leading to enhanced diagnostic assessability of surrounding tissues and target lesions. The study demonstrates the potential of iMAR and VMIs for PCDCT-guided interventions. These advancements could improve accuracy, safety, efficiency, and patient outcomes in clinical practice.
Keywords: artifact reduction; biopsy needle; computed tomography; interventional devices; iterative metal artifact reduction; phantom; photon-counting detector; photon-counting detector CT; radiofrequency ablation needle; virtual monoenergetic images.
Copyright © 2025 Wolters Kluwer Health, Inc. All rights reserved.
Conflict of interest statement
Conflicts of interest and sources of funding: S.F., B.S., and V.H. are employees of Siemens Healthineers (Erlangen, Germany). Y.C.L. was supported by a research grant from Siemens Healthineers (Erlangen, Germany). For the remaining authors, none were declared.
Similar articles
-
Metal Artifact Reduction in Photon-Counting Detector CT: Quantitative Evaluation of Artifact Reduction Techniques.Invest Radiol. 2024 Jun 1;59(6):442-449. doi: 10.1097/RLI.0000000000001036. Epub 2023 Oct 9. Invest Radiol. 2024. PMID: 37812482
-
Dental implant artifacts: Evaluation of photon counting CT-derived virtual monoenergetic images in combination with iterative metal artifact reduction algorithms.Eur J Radiol. 2025 Jun;187:112117. doi: 10.1016/j.ejrad.2025.112117. Epub 2025 Apr 14. Eur J Radiol. 2025. PMID: 40252281
-
Combining virtual monoenergetic imaging and iterative metal artifact reduction in first-generation photon-counting computed tomography of patients with dental implants.Eur Radiol. 2023 Nov;33(11):7818-7829. doi: 10.1007/s00330-023-09790-y. Epub 2023 Jun 7. Eur Radiol. 2023. PMID: 37284870 Free PMC article.
-
Impact of different metal artifact reduction techniques in photon-counting computed tomography head and neck scans in patients with dental hardware.Eur Radiol. 2024 Jun;34(6):3742-3749. doi: 10.1007/s00330-023-10430-8. Epub 2023 Nov 16. Eur Radiol. 2024. PMID: 37968474 Free PMC article.
-
Iterative metal artifact reduction on a clinical photon counting system-technical possibilities and reconstruction selection for optimal results dependent on the metal scenario.Phys Med Biol. 2022 Jun 7;67(11). doi: 10.1088/1361-6560/ac71f0. Phys Med Biol. 2022. PMID: 35594851
References
-
- Baum RA, Baum S. Interventional radiology: a half century of innovation. Radiology . 2014;273:S75–S91.
-
- Scheppers I, Wollschläger D. CT-guided biopsies and drainage. Radiologe . 2011;51:979–988; quiz 989-90, 990.
-
- Zhou W, Bartlett DJ, Diehn FE, et al. Reduction of metal artifacts and improvement in dose efficiency using photon-counting detector computed tomography and tin filtration. Invest Radiol . 2019;54:204–211.
-
- Lee M-J, Kim S, Lee SA, et al. Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT. Radiographics . 2007;27:791–803.
-
- Flohr T, Petersilka M, Henning A, et al. Photon-counting CT review. Phys Med . 2020;79:126–136.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical