Inability to adequately control antimicrobial agents on AutoMicrobic System Gram-Positive and Gram-Negative Susceptibility Cards
- PMID: 3980698
- PMCID: PMC271686
- DOI: 10.1128/jcm.21.3.454-456.1985
Inability to adequately control antimicrobial agents on AutoMicrobic System Gram-Positive and Gram-Negative Susceptibility Cards
Abstract
The Vitek Gram-Positive Susceptibility Card (GPS) and Gram-Negative General Susceptibility Plus Card (GSC Plus) were tested on the AutoMicrobic System (AMS) 50 times each with the recommended control organisms. Only 1 drug (chloramphenicol) of 11 on the GPS and 1 (gentamicin) of 10 on the GSC Plus could be adequately controlled, leaving unsubstantiated the results obtained with patient isolates on the remaining 19 antimicrobial agents.