Ultrasonographic evaluation of optic nerve sheath diameter in patients severe traumatic brain injury: a comparison with intraparenchymal pressure monitoring
- PMID: 39810071
- DOI: 10.1007/s10143-025-03202-z
Ultrasonographic evaluation of optic nerve sheath diameter in patients severe traumatic brain injury: a comparison with intraparenchymal pressure monitoring
Abstract
Objective: Increased intracranial pressure (ICP) can worsen the clinical condition of traumatic brain injury (TBI) patients. One non-invasive and easily bedside-performed technique to estimate ICP is ultrasonographic measurement of optic nerve sheath diameter (ONSD). This study aimed to analyze ONSD and correlate it with ICP values obtained by intraparenchymal monitoring to establish the ONSD threshold value for elevated ICP and reference range of ONSD in severe TBI patients.
Methods: Forty severe TBI patients (Glasgow Coma Scale Score ≤ 8) were included. Ultrasonographic measurement of ONSD was performed and compared with intraparenchymal ICP monitoring to assess their association and determine the ONSD threshold value. Exclusion criteria included individuals under eighteen years old, penetrating TBI, or direct ocular trauma.
Results: Fifty-three ONSD measurements were conducted in all patients. The mean ONSD value in the group with intracranial pressure < 20 mmHg was 5.4 mm ± 1.0, while in the group with intracranial pressure ≥ 20 mmHg, it was 6.4 mm ± 0.7 (p = 0.0026). A positive and statistically significant correlation, albeit weak (r = 0.33), was observed between ultrasonographic measurement of ONSD and intraparenchymal ICP monitoring. The statistical analysis of the ROC curve identified the best cut-off as 6.18 mm, with 77.8% sensitivity and 81.8% specificity.
Conclusion: Our results reveal a positive, albeit weak, correlation between ultrasonographic measurement of ONSD and intraparenchymal ICP monitoring, with an ONSD threshold value of 6.18 mm. Achieving only 77.8% sensitivity and considering the substantial variability between ONSD measurements (standard deviation at 1.0) might limit the reliability of ICP assessment based solely on ONSD measurements.
Keywords: Increased intracranial pressure; Intracranial pressure monitoring; Optic nerve sheath diameter; Traumatic brain injury; Ultrasonographic measurement.
© 2025. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.
Conflict of interest statement
Declarations. Competing Interests: The authors declare no competing interests. Ethics approval: This study was performed in line with the principles of the Declaration of Helsinki. Approval was granted by the Ethics Committee of the João XXIII Hospital (Date November 2017/No 78743317.8.0000.5119). Consent to participate: Written informed consent was obtained from family members or legal representatives of the participants.
Similar articles
-
Ultrasonographic measured optic nerve sheath diameter as an accurate and quick monitor for changes in intracranial pressure.J Neurosurg. 2015 Sep;123(3):743-7. doi: 10.3171/2014.10.JNS141197. Epub 2015 May 8. J Neurosurg. 2015. PMID: 25955869
-
Using Optic Nerve Sheath Diameter for Intracranial Pressure (ICP) Monitoring in Traumatic Brain Injury: A Scoping Review.Neurocrit Care. 2024 Jun;40(3):1193-1212. doi: 10.1007/s12028-023-01884-1. Epub 2023 Dec 19. Neurocrit Care. 2024. PMID: 38114797 Free PMC article.
-
Ultrasonographic Assessment of Optic Nerve Sheath Diameter as a Screening Tool for Intracranial Hypertension in Traumatic Brain Injury.World Neurosurg. 2024 Dec;192:e42-e48. doi: 10.1016/j.wneu.2024.08.111. Epub 2024 Aug 27. World Neurosurg. 2024. PMID: 39209251
-
Noninvasive Assessment of Intracranial Pressure: Deformability Index as an Adjunct to Optic Nerve Sheath Diameter to Increase Diagnostic Ability.Neurocrit Care. 2024 Oct;41(2):479-488. doi: 10.1007/s12028-024-01955-x. Epub 2024 Mar 6. Neurocrit Care. 2024. PMID: 38448744 Free PMC article.
-
Diagnostic accuracy of ultrasonographic optic nerve sheath diameter compared to invasive ICP monitoring for intracranial hypertension in pediatric patients.Pediatr Radiol. 2025 Jul;55(8):1669-1681. doi: 10.1007/s00247-025-06301-7. Epub 2025 Jun 17. Pediatr Radiol. 2025. PMID: 40526116
Cited by
-
Optic nerve sheath diameter measurement in healthy South Iranian adults from a cross-sectional ultrasonographic study.Sci Rep. 2025 May 19;15(1):17404. doi: 10.1038/s41598-025-01891-6. Sci Rep. 2025. PMID: 40389493 Free PMC article.
-
Bedside Ultrasonographic Measurement of Optic Nerve Sheath Diameter for Assessing Increased Intracranial Pressure: An Observational Study.Cureus. 2025 Jun 16;17(6):e86163. doi: 10.7759/cureus.86163. eCollection 2025 Jun. Cureus. 2025. PMID: 40677488 Free PMC article.
References
-
- Carney N, Totten AM, O’Reilly C, Ullman JS, Hawryluk GW, Bell MJ, Bratton SL, Chesnut R, Harris OA, Kissoon N, Rubiano AM, Shutter L, Tasker RC, Vavilala MS, Wilberger J, Wright DW, Ghajar J. Guidelines for the management of severe traumatic brain injury, forth edition. 2017;80(1): 6-15
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical