Alveolar pressure swings during high frequency ventilation in rabbits
- PMID: 3982871
- DOI: 10.1203/00006450-198502000-00002
Alveolar pressure swings during high frequency ventilation in rabbits
Abstract
In order to understand the pressures and volumes delivered during ventilation with the Emerson flow-interrupting high frequency ventilator, we have measured dynamic and mean pressures at the airway opening, trachea, and alveoli as well as delivered volumes in vivo in closed-chest adult rabbits ventilated at rates of 2-37.5 Hz. To measure alveolar pressure we opened the chest, glued an alveolar capsule to the visceral pleural surface, punctured the pleura and lung surface through the capsule, inserted a pressure transducer into the capsule and closed the chest. We measured delivered volume with a pressure plethysmograph. Alveolar pressure swings fell with increasing frequency, as did delivered volume. Alveolar pressure swings were always lower than those at the airway opening or in the trachea. Removal of the rib cage resulted in a decrease in alveolar pressure swings although muscle paralysis had little effect. Mean pressures were equal at the airway opening, trachea, and alveoli at all frequencies. These results suggest that using the high frequency ventilator tested, to minimize alveolar barotrauma one should choose the highest frequency where gas exchange is adequate.
Similar articles
-
Elevated lung volume and alveolar pressure during jet ventilation of rabbits.Am Rev Respir Dis. 1985 Jan;131(1):134-8. doi: 10.1164/arrd.1985.131.1.134. Am Rev Respir Dis. 1985. PMID: 3966700
-
Proximal, tracheal, and alveolar pressures during high-frequency oscillatory ventilation in a normal rabbit model.Pediatr Res. 1990 Oct;28(4):367-73. doi: 10.1203/00006450-199010000-00013. Pediatr Res. 1990. PMID: 2235135
-
Alveolar pressure magnitude and asynchrony during high-frequency oscillations of excised rabbit lungs.Am Rev Respir Dis. 1985 Aug;132(2):343-9. doi: 10.1164/arrd.1985.132.2.343. Am Rev Respir Dis. 1985. PMID: 4026057
-
Mean airway pressure and mean alveolar pressure during high-frequency jet ventilation in rabbits.J Appl Physiol (1985). 1986 Aug;61(2):456-63. doi: 10.1152/jappl.1986.61.2.456. J Appl Physiol (1985). 1986. PMID: 3745038
-
Setting the frequency-tidal volume pattern.Respir Care. 2002 Mar;47(3):266-74; discussion 274-8. Respir Care. 2002. PMID: 11874606 Review.
Cited by
-
[High-frequency oscillatory ventilation. Ventilation procedure for adults with acute lung failure].Anaesthesist. 2007 May;56(5):485-90. doi: 10.1007/s00101-007-1167-z. Anaesthesist. 2007. PMID: 17375267 Review. German.
-
Prevention of bronchopulmonary dysplasia.Semin Fetal Neonatal Med. 2009 Dec;14(6):374-82. doi: 10.1016/j.siny.2009.08.002. Epub 2009 Sep 6. Semin Fetal Neonatal Med. 2009. PMID: 19736053 Free PMC article.
-
Scanographic comparison of high frequency oscillation with versus without tracheal gas insufflation in acute respiratory distress syndrome.Intensive Care Med. 2011 Jun;37(6):990-9. doi: 10.1007/s00134-011-2162-z. Epub 2011 Mar 3. Intensive Care Med. 2011. PMID: 21369813
-
Effect of the I/E ratio on CO2 removal during high-frequency oscillatory ventilation with volume guarantee in a neonatal animal model of RDS.Eur J Pediatr. 2016 Oct;175(10):1343-51. doi: 10.1007/s00431-016-2770-2. Epub 2016 Sep 5. Eur J Pediatr. 2016. PMID: 27595847
-
Impaired response to deep inspiration in obesity.J Appl Physiol (1985). 2011 Sep;111(3):726-34. doi: 10.1152/japplphysiol.01155.2010. Epub 2011 Jun 23. J Appl Physiol (1985). 2011. PMID: 21700888 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources