. 2025 Feb:130:104904.
doi: 10.1016/j.ejmp.2025.104904. Epub 2025 Jan 21.

Characterization and evaluation methods of fused deposition modeling and stereolithography additive manufacturing for clinical linear accelerator photon and electron radiotherapy applications

Affiliations
Free article

Characterization and evaluation methods of fused deposition modeling and stereolithography additive manufacturing for clinical linear accelerator photon and electron radiotherapy applications

John Paul O Bustillo et al. Phys Med. 2025 Feb.
Free article

Abstract

Purpose: To propose comprehensive characterization methods of additive manufacturing (AM) materials for MV photon and MeV electron radiotherapy.

Methodology: This study investigated 15 AM materials using CT machines. Geometrical accuracy, tissue-equivalence, uniformity, and fabrication parameters were considered. Selected soft tissue equivalent filaments were used to fabricate slab phantoms and compared with water equivalent RW3 phantom by delivering planar 6 & 10 MV photons and 6, 9, 12, 15, & 18 MeV electrons. Finally, a 3D printed CT-Electron Density characterization phantom was fabricated.

Results: Materials used to print test objects can simulate tissues from adipose (relative electron density, ρe=0.72) up to near inner bone-equivalent (ρe=1.08). Lower densities such as breast and lung can be simulated using infills from 90 % down to 30 %, respectively. The gyroid infill pattern shows the lowest CT number variation and is recommended for low infill percentage printing. CT number uniformity can be observed from 40 % up to 100 % infill, while printing orientation does not significantly affect the CT number. The measured doses using the 3D printed phantoms show to have good agreement with TPS calculated dose for photon (< 1 % difference) and electron (< 5 % difference). Varying the printed slab thicknesses shows very similar response (< 3 % difference) compared with RW3 slabs except for 6 MeV electrons. Lastly, the fabricated CT-ED phantom generally matches the lung- up to the soft tissue- equivalence.

Conclusion: The proposed methods give the outline for characterization of AM materials as tissue-equivalent substitute. Printing parameters affect the radiological quality of 3D-printed object.

Keywords: 3D printing; Additive manufacturing; Dosimetry; Fused deposition modeling; Phantom; Protocol; Radiotherapy; Stereolithography.

PubMed Disclaimer

Conflict of interest statement

Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.