Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2025 Feb;85(1):11-19.
doi: 10.1080/00365513.2025.2456917. Epub 2025 Jan 25.

Glucometer versus analyzer: comparable results with negligible clinical risk

Affiliations
Comparative Study

Glucometer versus analyzer: comparable results with negligible clinical risk

Alper Gümüş et al. Scand J Clin Lab Invest. 2025 Feb.

Abstract

This study assessed the reliability of Roche Accu-Chek Inform II glucometers in a real-world setting. A retrospective analysis was conducted on 6,695 paired results. Capillary samples were tested using Roche Accu-Chek Inform II glucometers, while venous samples were analyzed using Roche Cobas c503/702 analyzers. Compliance was assessed using modified criteria based on the ISO 15197 guideline and the CLSI EP09-A3 guideline using Passing-Bablok regression analysis, Bland-Altman plots, and Surveillance Error Grid (SEG) analysis. The overall compliance of glucometer results within ±15% or 0.83 mmol/L (15 mg/dL) of the reference method was 81.5%, below the acceptance criterion of 94.6%. SEG analysis showed that 90.3% of the paired results fell within the No-risk zone, with less than 0.001% in the moderate/lower-risk zone. The Emergency Department results indicated 87.8% overall compliance and 92.2% of pairs falling in the No-risk zone. Based on the regression analysis, the glucometer results showed a positive constant bias of nearly 0.33 mmol/L (6 mg/dL). The Bland-Altman plots showed a positive mean difference of 0.43 mmol/L for results ≤5.55 mmol/L (≤100 mg/dL) and a positive mean percentage difference of 3.77% for results >5.55 mmol/L (>100 mg/dL), within the permissible deviation. The compliance values ranged from 76.0% to 90.3% for clinical concentration groups, with the highest compliance found between >16.65-22.20 mmol/L (>300-400 mg/dL). The Accu-Chek Inform II glucometers demonstrated in real-world reliability, with most results falling within acceptable risk categories. However, compliance still needs improvement, so manufacturers should assess opportunities for advancement.

Keywords: Blood glucose; blood glucose analysis; hospitals; post-marketing product surveillance; surveillance error grid.

PubMed Disclaimer

References

Publication types

LinkOut - more resources