High-precision localization of radiation isocenter using Winston-Lutz test: Impact of collimator angle, phantom position, and field size
- PMID: 39904583
- PMCID: PMC11969077
- DOI: 10.1002/acm2.70000
High-precision localization of radiation isocenter using Winston-Lutz test: Impact of collimator angle, phantom position, and field size
Abstract
Purpose: This study aimed to evaluate the impact of collimator angle, ball bearing (BB) phantom position, and field size on the accuracy of Winston-Lutz (WL) test-derived radiation isocenters.
Methods: WL tests were performed on four TrueBeam linear accelerators. Fifty-six images (eight gantry angles multiplied by seven collimator angles) were acquired for each WL test. Images with different sets of collimator angles were used to compute the radiation isocenters. The resulting radiation isocenters were correlated with the collimator angles. Then, the BB position and radiation field size were varied for the subsequent WL tests. The calculated BB shifts were compared with the known shifts, and the radiation isocenters were compared between different field sizes.
Results: The use of a single collimator angle led to errors of as much as 0.4 mm in the calculated radiation isocenters. Systematic differences were observed between the radiation isocenters derived with collimator angle 0° and those derived with 90° and/or 270°. A commonly used opposing collimator angle pair, 90° and 270°, resulted in a vertical 0.1 mm offset of the radiation isocenters toward the ceiling. Oblique opposite or mixed collimator angles yielded radiation isocenter errors less than 0.1 mm. The BB shifts derived from WL tests were less than 0.1 mm from the known shifts. The radiation isocenters varied by less than 0.1 mm between field sizes ranging from 2 × 2 cm2 to 20 × 20 cm2.
Conclusions: Oblique opposing collimator angle pairs should be considered to minimize errors in localizing radiation isocenters. Uncertainty in BB positioning could be eliminated if the BB is used as a static reference point in space. The field size had no significant effect on the radiation isocenters. With careful design of WL test parameters and image processing, it is possible to achieve a precision of 0.1 mm in localizing radiation isocenters using WL tests.
Keywords: Winston‐Lutz test; quality assurance; radiation isocenter.
© 2025 The Author(s). Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.
Conflict of interest statement
The author declares no conflicts of interest.
Figures





Similar articles
-
On the selection of gantry and collimator angles for isocenter localization using Winston-Lutz tests.J Appl Clin Med Phys. 2016 Jan 8;17(1):167-178. doi: 10.1120/jacmp.v17i1.5792. J Appl Clin Med Phys. 2016. PMID: 26894350 Free PMC article.
-
Off-iso Winston-Lutz test on seven linear accelerators.J Appl Clin Med Phys. 2024 Oct;25(10):e14470. doi: 10.1002/acm2.14470. Epub 2024 Jul 23. J Appl Clin Med Phys. 2024. PMID: 39042435 Free PMC article.
-
Efficient quality assurance method with automated data acquisition of a single phantom setup to determine radiation and imaging isocenter congruence.J Appl Clin Med Phys. 2019 Oct;20(10):127-133. doi: 10.1002/acm2.12723. Epub 2019 Sep 19. J Appl Clin Med Phys. 2019. PMID: 31535781 Free PMC article.
-
A novel phantom and procedure providing submillimeter accuracy in daily QA tests of accelerators used for stereotactic radiosurgery*.J Appl Clin Med Phys. 2016 Jul 8;17(4):246-253. doi: 10.1120/jacmp.v17i4.6295. J Appl Clin Med Phys. 2016. PMID: 27455506 Free PMC article.
-
Alignment of multiradiation isocenters for megavoltage photon beam.J Appl Clin Med Phys. 2015 Nov 8;16(6):314-324. doi: 10.1120/jacmp.v16i6.5733. J Appl Clin Med Phys. 2015. PMID: 26699586 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical