Anatomic analysis of removed prosthetic heart valves: causes of failure of 33 mechanical valves and 58 bioprostheses, 1980 to 1983
- PMID: 3997132
- DOI: 10.1016/s0046-8177(85)80103-9
Anatomic analysis of removed prosthetic heart valves: causes of failure of 33 mechanical valves and 58 bioprostheses, 1980 to 1983
Abstract
The details of heart valve prosthesis-associated problems are not widely known. This study investigated the etiologies of the failures of 91 valves, 33 mechanical prostheses and 58 bioprostheses, obtained at reoperation (83) or autopsy (eight) at the Brigham and Women's Hospital during the 42-month period from mid- 1980 through 1983, one to 264 months (mean, 72 months) after valve replacement. Analysis was by gross, histologic, radiographic, and microbiologic examination, as well as review of clinical records. Overall causes of failure included paravalvular leak (15 per cent), thrombosis (7 per cent), tissue overgrowth (8 per cent), degeneration or mechanical failure (43 per cent), and endocarditis (19 per cent). Endocarditis and paravalvular leak were equally frequent with mechanical prostheses and bioprostheses. In addition, thrombosis (18 per cent), tissue overgrowth (21 per cent), and structural failure (12 per cent) were all important failure modes for mechanical prostheses. Sterile degeneration was the overwhelming cause of failure for bioprostheses, accounting for the failure of 35 of 58 (60 per cent) of those recovered. Sterile degeneration took several forms: calcification, with or without cuspal tears (27 cases, 47 per cent of bioprostheses; mean, 77 months, range, 44 to 108 months) and cuspal defects without calcification (eight cases, 14 per cent; mean, 59 months, range, eight to 122 months). In general, calcification increased with time after implantation, but the propensity for the mineralization of bioprostheses varied widely among patients. Four torn valves that had been in place for more than six years had radiographically undetectable calcific deposits. The results of this study indicate that paravalvular leak and endocarditis are frequent causes of failure for all valve types. No clear failure mode predominates with mechanical valve prostheses, although some designs have specific inherent limitations. In contrast, degeneration, especially that related to mineralization, is the most important cause of the late failure of contemporary bioprostheses.
Similar articles
-
Pathology of the Pericarbon bovine pericardial xenograft implanted in humans.J Heart Valve Dis. 1998 Mar;7(2):180-9. J Heart Valve Dis. 1998. PMID: 9587859
-
Causes of failure and pathologic findings in surgically removed Ionescu-Shiley standard bovine pericardial heart valve bioprostheses: emphasis on progressive structural deterioration.Circulation. 1987 Sep;76(3):618-27. doi: 10.1161/01.cir.76.3.618. Circulation. 1987. PMID: 3621522
-
[Structural changes in valve prostheses].Cardiologia. 1991 Dec;36(12 Suppl 1):91-6. Cardiologia. 1991. PMID: 1841811 Review. Italian.
-
Morphologic findings in explanted Hancock II porcine bioprostheses.J Heart Valve Dis. 1999 Jan;8(1):4-15. J Heart Valve Dis. 1999. PMID: 10096476
-
[Dysfunctions of heart valve prostheses and their surgical treatment].Schweiz Med Wochenschr. 1987 Oct 24;117(43):1665-70. Schweiz Med Wochenschr. 1987. PMID: 3321421 Review. French.
Cited by
-
An analysis of autopsy findings in 108 patients who died after valve replacement.Virchows Arch A Pathol Anat Histopathol. 1993;422(5):397-403. doi: 10.1007/BF01605459. Virchows Arch A Pathol Anat Histopathol. 1993. PMID: 8322455
-
Clinical problems in patients with prosthetic heart valves.J R Coll Physicians Lond. 1990 Jul;24(3):167-72. J R Coll Physicians Lond. 1990. PMID: 2213670 Free PMC article. Review.
-
Pre- and post-surgical evaluation of the inflammatory response in patients with aortic stenosis treated with different types of prosthesis.BMC Cardiovasc Disord. 2017 Apr 14;17(1):100. doi: 10.1186/s12872-017-0526-1. BMC Cardiovasc Disord. 2017. PMID: 28410571 Free PMC article.
-
A new method for selecting calf pericardium for use in cardiac bioprostheses on the basis of morphological and mechanical criteria.J Mater Sci Mater Med. 2001 Aug;12(8):665-71. doi: 10.1023/a:1011252022303. J Mater Sci Mater Med. 2001. PMID: 15348235
-
[Mechanical versus biological heart valves].Z Kardiol. 2001 Dec;90(Suppl 6):39-47. doi: 10.1007/s003920170007. Z Kardiol. 2001. PMID: 24445787 German.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources