Comparison of a high performance liquid chromatographic and an enzyme immunoassay technique for quantitation of theophylline in serum
- PMID: 399811
- DOI: 10.1097/00007691-197901000-00007
Comparison of a high performance liquid chromatographic and an enzyme immunoassay technique for quantitation of theophylline in serum
Abstract
High performance liquid chromatograhic (HPLC) and enzyme immunoassay (EMIT) methods for quantitating theophylline concentrations in serum were evaluated. For both methods, standard curves were linear over the therapeutic range of serum concentrations. Precision was acceptable, with coefficients of variation being less than 9%. The HPLC assay was slightly, but not significantly, more precise. No interference was noted in either method by caffeine, theobromine, diphylline, 8-chlorotheophylline, or by metabolites such as 1-methylxanthine, and 3-methylxanthine, or by potentially interfering compounds such as urea and uric acid. The two methods correlated well, with a correlation coefficient of 0.98%. The enzyme immunoassay was superior in terms of lower costs, greater ease of performance, and potential for automation.
Similar articles
-
Comparison of homogeneous enzyme immunoassay and high-pressure liquid chromatography for the determination of theophylline concentration in serum.Am Rev Respir Dis. 1978 Jun;117(6):1135-8. doi: 10.1164/arrd.1978.117.6.1135. Am Rev Respir Dis. 1978. PMID: 352208
-
Concurrent measurement of theophylline and caffeine in neonates by an interference-free liquid-chromatographic method.Clin Chem. 1983 Nov;29(11):1934-6. Clin Chem. 1983. PMID: 6354512
-
Evaluation of caffeine plasma levels by an automated enzyme immunoassay (EMIT) in comparison with a high-performance liquid chromatographic method.Ther Drug Monit. 1984;6(3):348-54. doi: 10.1097/00007691-198409000-00016. Ther Drug Monit. 1984. PMID: 6390797
-
Evaluation of a rapid high-performance liquid chromatographic method for plasma theophylline analysis and its use in therapeutic drug monitoring.Ther Drug Monit. 1984;6(3):322-7. doi: 10.1097/00007691-198409000-00012. Ther Drug Monit. 1984. PMID: 6390795
-
Methods of theophylline assay and therapeutic monitoring of this drug.Ann Biol Clin (Paris). 1990;48(5):287-93. Ann Biol Clin (Paris). 1990. PMID: 2195925 Review.
Cited by
-
Theophylline.Clin Rev Allergy. 1983 Mar;1(1):73-85. doi: 10.1007/BF02991318. Clin Rev Allergy. 1983. PMID: 6367926 Review. No abstract available.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources