Case Reports
. 2024 Dec 26;9(2):60-64.
doi: 10.1016/j.case.2024.10.003. eCollection 2025 Feb.

Lethal Myocardial Infarction in a Neonate During Extracorporeal Membrane Oxygenation

Affiliations
Case Reports

Lethal Myocardial Infarction in a Neonate During Extracorporeal Membrane Oxygenation

Numra A Aleem et al. CASE (Phila). .
No abstract available

Keywords: Coronary artery thrombus; Extracorporeal membrane oxygenation; Myocardial infarction; Targeted neonatal echocardiography.

PubMed Disclaimer

Figures

None
Graphical abstract
Figure 1
Figure 1
Chest radiograph, anterior-posterior projection, demonstrates the VA-ECMO cannula positions (red circles) with the arterial cannula tip terminated at T6, which was deeper than its location at insertion.
Figure 2
Figure 2
Two-dimensional transthoracic echocardiography, parasternal long-axis view, demonstrates a clot extending from the proximal ascending aorta into the LV cavity.
Figure 3
Figure 3
Two-dimensional transthoracic echocardiography, parasternal long-axis view, demonstrates the arterial cannula extending deep into the proximal ascending aorta.
Figure 4
Figure 4
Two-dimensional transthoracic echocardiography with color-flow Doppler (reduced Nyquist limit), parasternal short-axis view, demonstrates no evidence of flow in the LMCA.
Figure 5
Figure 5
Two-dimensional transthoracic echocardiography, apical 5-chamber view in systole (left) and diastole (right), demonstrates that the aortic valve (red circle) was not opening at any point during the cardiac cycle.
Figure 6
Figure 6
Electrocardiographic rhythm strip recording (precordial lead V2) during clinical deterioration demonstrates marked ST-segment elevation.

References

    1. MacLaren G., Brodie D., Lorusso R., Peek G., Thiagarajan R., Vercaemst L. Extracorporeal Life Support: The ELSO Red Book. Extracorporeal Life Support Organization; Ann Arbor: 2022.
    1. Kamdar A., Rintoul N., Raffini L. Anticoagulation in neonatal ECMO. Semin Perinatol. 2018;42:122–128. - PubMed
    1. Ozment C.P., Scott B.L., Bembea M.M., Spinella P.C., Lu P.E.S.P.A., Org E.L.S. Anticoagulation and Transfusion management during neonatal and pediatric extracorporeal membrane oxygenation: a survey of medical directors in the United States∗. Pediatr Crit Care Me. 2021;22:530–541. - PubMed
    1. Dalton H.J., Reeder R., Garcia-Filion P., Holubkov R., Berg R.A., Zuppa A., et al. Factors associated with bleeding and thrombosis in children receiving extracorporeal membrane oxygenation. Am J Respir Crit Care Med. 2017;196:762–771. - PMC - PubMed
    1. Williams B., Bernstein W. Review of venoarterial extracorporeal membrane oxygenation and development of intracardiac thrombosis in adult cardiothoracic patients. J Extra Corpor Technol. 2016;48:162–167. - PMC - PubMed

Publication types