Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Multicenter Study
. 2025 Aug 1;46(7):789-795.
doi: 10.1097/MAO.0000000000004438. Epub 2025 Feb 25.

A Multi-Institutional Analysis of Device Failures in a Recent Cochlear Implant Series

Affiliations
Multicenter Study

A Multi-Institutional Analysis of Device Failures in a Recent Cochlear Implant Series

Michael H Freeman et al. Otol Neurotol. .

Abstract

Objective: To assess Advanced Bionics (AB) HiRes Ultra and Ultra 3D initial version (V1) cochlear implant electrode failures over time at four large cochlear implant programs.

Study design: Retrospective cohort.

Setting: Four tertiary referral centers.

Patients: Patients receiving AB HiRes Ultra and Ultra 3D initial version (V1) devices as of December 31, 2022.

Main outcome measures: Failure rate, revision surgery, speech recognition scores.

Results: Over a span of 7 years, 205 (42.6%) of 483 implanted V1 devices have failed. Device failure rate varied across institutions from 32 to 67%. Of the 205 detected failures, 163 (79%) have undergone revision surgery, with 94% of revisions being performed with Advanced Bionics devices. Average time from implantation to diagnosis of device failure was 2.7 ± 1.2 years. Most patients (89.8%) with available testing were matched or exceeded their best prefailure speech performance after implant revision. After revision, patients had an average CNC score improvement of 24.7 percentage points over their most recent pre-revision scores and demonstrated average datalogging of 12.7 ± 4.1 h/d at most recent evaluation.

Conclusions: Comparison of patients across multiple high-volume implant centers confirms the presence of ongoing device failures. There is variability across institutions in the rate of revision surgery once a patient is diagnosed with a V1 device failure, as well as in the rate of device failure detection. Inter-institutional variability in failure rates may be explained by the variation in the routine use of electrical field imaging. Reimplantation with a new device typically results in a return to prefailure peak performance.

Professional practice gap educational need: To our knowledge, a comparison of AB HiRes (V1) device failures across multiple institutions has not been conducted.

Learning objective: To identify device failure rate across multiple institutions with different testing protocols.

Desired result: Providers will have an improved understanding of the trajectory of device failures for HiRes (V1) devices over time.

Level of evidence: Level IV-Historical cohort or case-controlled studies.

Indicate irb or iacuc: IRB #230017; IRB #22-000183; IRB #STU-032018-085; IRB #H-49479.

Keywords: Cochlear implants; Device failure; Surgical outcomes.

PubMed Disclaimer

References

    1. Carlson ML, Driscoll CLW, Gifford RH, McMenomey SO. Cochlear implantation: Current and future device options. Otolaryngol Clin North Am 2012;45:221–48.
    1. Carlson ML. Cochlear implantation in adults. N Engl J Med 2020;382:1531–42.
    1. Lane C, Zimmerman K, Agrawal S, Parnes L. Cochlear implant failures and reimplantation: A 30-year analysis and literature review. Laryngoscope 2020;130:782–9.
    1. Kim SY, Kim MB, Chung WH, et al. Evaluating reasons for revision surgery and device failure rates in patients who underwent cochlear implantation surgery. JAMA Otolaryngol Head Neck Surg 2020;146:414–20.
    1. Hildrew DM, Molony TB. Nucleus N5 CI500 series implant recall: hard failure rate at a major cochlear implantation center. Laryngoscope 2013;123:2829–33.

Publication types