Two-Dimensional Preoperative Digital Templating is Less Accurate When Using a Collared Triple Taper Stem Versus a Single Taper Design
- PMID: 40130235
- PMCID: PMC11932654
- DOI: 10.1016/j.artd.2025.101658
Two-Dimensional Preoperative Digital Templating is Less Accurate When Using a Collared Triple Taper Stem Versus a Single Taper Design
Abstract
Background: Collared triple taper stems (CTTS) and single taper stems (STS) have an excellent performance in primary total hip arthroplasty (THA). While 2-dimensional radiographic templating is accurate for STS, data are lacking for CTTS. We hypothesize that CTTS' more anatomical design in the anteroposterior dimension may lead to inaccurate templating sizing. This study compared templating accuracy of CTTS to a predecessor STS in THA patients.
Methods: 106 THA performed with CTTS were compared to 106 THA performed with STS by 2 high-volume surgeons. The stems chosen for comparison were manufactured by the same company, use the same templating software, shared an identical medial-lateral profile, and offered the same size range. The ability of digital templating to predict final implant size was evaluated.
Results: Template to stem accuracy was 36.8% for CTTS and 49.1% for STS (P = .07). Accuracy within 1 size was 88.7% for CTTS versus 95.2% for STS (P = .1). CTTS was implanted using a smaller size compared to the template twice as frequently as STS (43.4% vs 20.8%; P < .01). CTTS was 3.7 times more likely to have implants 2 or more sizes under the template compared to STS (10.4% vs 2.8%; P = .02). In logistic regression, the only predictor of implant 2+ sizes under the template was type of stem (P = .04).
Conclusions: The accuracy of conventional templating for CTTS is lower than the predecessor STS, with the template often suggesting a larger size. Bi-planar or 3-dimensional preoperative templating could potentially be a more accurate technique, especially during the initial learning curve with these stems.
Keywords: Templating; Total hip arthroplasty; Triple taper stems.
© 2025 The Authors.
Figures


Similar articles
-
Digital Templating of THA Using PACS and an iPhone or iPad is as Accurate as Commercial Digital Templating Software.Clin Orthop Relat Res. 2023 Jun 1;481(6):1104-1113. doi: 10.1097/CORR.0000000000002474. Epub 2022 Nov 18. Clin Orthop Relat Res. 2023. PMID: 36730564 Free PMC article.
-
Do the sizes of total hip arthroplasty implants match between 3D planning software and 2D templating?Orthop Traumatol Surg Res. 2024 Apr;110(2):103744. doi: 10.1016/j.otsr.2023.103744. Epub 2023 Nov 3. Orthop Traumatol Surg Res. 2024. PMID: 37923177
-
Determining the accuracy of preoperative total hip replacement 2D templating using the mediCAD® software.J Orthop Surg Res. 2022 Apr 10;17(1):222. doi: 10.1186/s13018-022-03086-5. J Orthop Surg Res. 2022. PMID: 35399090 Free PMC article.
-
How to perform correct templating in total hip replacement.Musculoskelet Surg. 2023 Mar;107(1):19-28. doi: 10.1007/s12306-023-00772-3. Epub 2023 Jan 11. Musculoskelet Surg. 2023. PMID: 36630067 Review.
-
Comparative outcomes between collared versus collarless and short versus long stem of direct anterior approach total hip arthroplasty: a systematic review and indirect meta-analysis.Eur J Orthop Surg Traumatol. 2019 Dec;29(8):1693-1704. doi: 10.1007/s00590-019-02516-1. Epub 2019 Jul 30. Eur J Orthop Surg Traumatol. 2019. PMID: 31363848
References
-
- Panichkul P., Bavonratanavech S., Arirachakaran A., Kongtharvonskul J. Comparative outcomes between collared versus collarless and short versus long stem of direct anterior approach total hip arthroplasty: a systematic review and indirect meta-analysis. Eur J Orthop Surg Traumatol. 2019;29:1693–1704. doi: 10.1007/s00590-019-02516-1. - DOI - PubMed