Evaluation of silicone elastomer catheters for long-term intravenous chemotherapy
- PMID: 4015268
Evaluation of silicone elastomer catheters for long-term intravenous chemotherapy
Abstract
We evaluated the efficacy and the complications of 65 silicone elastomer catheters inserted percutaneously for long-term venous access for administration of chemotherapy, antibiotics, and blood products in patients with metastatic cancer. Treatments were administered either in the hospital or in the outpatient clinic, using a portable infusion pump. The median indwelling time of catheters was 238 days (range, two to 521). The projected duration of catheter function, when the electively removed catheters were censured, was 310 days. Twenty-three catheters were removed because of malfunction, while the remaining either were discontinued electively (20) or were functioning at the conclusion of the study (22). The problems necessitating removal of 23 catheters were inadvertent dislodgement from loose sutures (eight), mechanical damage to the catheters (four), sepsis (four), phlebitis (four), intraluminal blockage with a clot (two), and cellulitis (one). We conclude that silicone elastomer catheters are safe and reliable for extended venous access for cancer chemotherapy. They are easy to insert and remove and can be replaced with a guide wire without requiring surgical intervention.
Similar articles
-
Long-term indwelling silicone catheters: a simple, safe method for venous access.Can J Surg. 1987 May;30(3):181-4. Can J Surg. 1987. PMID: 3580975
-
Long-term intravenous therapy with peripherally inserted silicone elastomer central venous catheters in patients with malignant diseases.Cancer. 1979 May;43(5):1937-43. doi: 10.1002/1097-0142(197905)43:5<1937::aid-cncr2820430554>3.0.co;2-t. Cancer. 1979. PMID: 109182
-
Low infection rate and long durability of nontunneled silastic catheters. A safe and cost-effective alternative for long-term venous access.Arch Intern Med. 1993 Aug 9;153(15):1791-6. Arch Intern Med. 1993. PMID: 8392831
-
[Implantable catheter systems].Praxis (Bern 1994). 1998 Sep 2;87(36):1135-9. Praxis (Bern 1994). 1998. PMID: 9782741 Review. German.
-
Large bore catheters with surface treatments versus untreated catheters for vascular access in hemodialysis.Artif Organs. 2004 Jul;28(7):604-10. doi: 10.1111/j.1525-1594.2004.07365.x. Artif Organs. 2004. PMID: 15209851 Review.
Cited by
-
Peripherally inserted central venous lines versus central lines in surgical newborns--a comparison.Indian J Pediatr. 2010 Feb;77(2):171-4. doi: 10.1007/s12098-009-0291-y. Epub 2010 Jan 20. Indian J Pediatr. 2010. PMID: 20091380
-
Peripherally inserted central venous catheters. Low-risk alternatives for ongoing venous access.West J Med. 1994 Jan;160(1):25-30. West J Med. 1994. PMID: 8128698 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical