Artefactual ST segment abnormalities due to electrocardiograph design
- PMID: 4015923
- PMCID: PMC481865
- DOI: 10.1136/hrt.54.2.121
Artefactual ST segment abnormalities due to electrocardiograph design
Abstract
Specifications recommended for electrocardiographs omit any reference to phase characteristics and place undue emphasis on an extended amplitude response. This does not, however, ensure faithful reproduction of the electrocardiogram and may result in less effective attenuation of unwanted noise than can be achieved with modern filters. The separate effects of phase and amplitude response on electrocardiographic signals were compared, and it was shown that distortion of the electrocardiogram by equipment with an inadequate phase response may cause changes resembling those seen in ischaemic heart disease. Case reports are presented in which distortion produced by commercial equipment impeded diagnosis. Specifications recommended as standard for electrocardiographs should be modified to include phase response and allow greater attenuation of frequencies below 0.5 Hz.
Similar articles
-
Effects of electrocardiograph frequency filters on P-QRS-T amplitudes of the feline electrocardiogram.Am J Vet Res. 1995 Nov;56(11):1534-40. Am J Vet Res. 1995. PMID: 8585669
-
[ST-segment analysis in long-term ECG: amplitude and phase response of various systems in comparison with standard ECG and their effect on true original reproduction of ST segment depression].Z Kardiol. 1989 Jan;78(1):14-22. Z Kardiol. 1989. PMID: 2922953 German.
-
Simulation of anteroseptal myocardial infarction by electrocardiographic filters.J Electrocardiol. 2006 Jul;39(3):253-8. doi: 10.1016/j.jelectrocard.2005.11.001. Epub 2006 Feb 28. J Electrocardiol. 2006. PMID: 16777511
-
What is inside the electrocardiograph?J Electrocardiol. 2008 Jan-Feb;41(1):8-14. doi: 10.1016/j.jelectrocard.2007.08.059. J Electrocardiol. 2008. PMID: 18191652 Review.
-
Equipment-related electrocardiographic artifacts: causes, characteristics, consequences, and correction.Anesthesiology. 2008 Jan;108(1):138-48. doi: 10.1097/01.anes.0000296537.62905.25. Anesthesiology. 2008. PMID: 18156892 Review.
Cited by
-
ANMCO/AIIC/SIT Consensus Information Document: definition, precision, and suitability of electrocardiographic signals of electrocardiographs, ergometry, Holter electrocardiogram, telemetry, and bedside monitoring systems.Eur Heart J Suppl. 2017 May;19(Suppl D):D190-D211. doi: 10.1093/eurheartj/sux031. Epub 2017 May 2. Eur Heart J Suppl. 2017. PMID: 28751842 Free PMC article.
-
Continuous ambulatory pulmonary artery pressure monitoring. A new method using a transducer tipped catheter and a simple recording system.Br Heart J. 1986 Apr;55(4):336-43. doi: 10.1136/hrt.55.4.336. Br Heart J. 1986. PMID: 3964499 Free PMC article.
-
High-Bandpass Filters in Electrocardiography: Source of Error in the Interpretation of the ST Segment.ISRN Cardiol. 2012;2012:706217. doi: 10.5402/2012/706217. Epub 2012 Jun 21. ISRN Cardiol. 2012. PMID: 22778996 Free PMC article.
-
Do we need a new standard for electrocardiographs?Br Heart J. 1985 Aug;54(2):119-20. doi: 10.1136/hrt.54.2.119. Br Heart J. 1985. PMID: 4015922 Free PMC article. No abstract available.
-
Detection of intraoperative myocardial ischaemia--a comparison among electrocardiographic, myocardial metabolic, and haemodynamic measurements in patients with reduced ventricular function.Can J Anaesth. 1995 Jun;42(6):487-94. doi: 10.1007/BF03011686. Can J Anaesth. 1995. PMID: 7628028
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources