Clinical evaluation of three rapid methods for the detection of significant bacteriuria
- PMID: 4019744
- PMCID: PMC268322
- DOI: 10.1128/jcm.22.1.62-67.1985
Clinical evaluation of three rapid methods for the detection of significant bacteriuria
Abstract
Random urine specimens (848) were screened for significant bacteriuria by using the 30-min Lumac (3M, St. Paul, Minn.), the 2-min Bac-T-Screen (Marion Laboratories, Inc., Kansas City, Mo.), and the 13-h AutoMicrobic system (AMS) urine identification card (Vitek Systems, Inc., Hazelwood, Mo.). MacConkey and 5% sheep blood agar plates were inoculated with a 10(-4) dilution of urine and used for the reference method. Bac-T-Screen results were uninterpretable for 9.1% of the specimens owing to either urine sample pigmentation (5.3%) or clogging of the filter (3.8%). Screen-negative urine specimens made up 49.6, 57.2, and 48.5% of the total number of specimens evaluated with AMS, Lumac, and Bac-T-Screen, respectively. False-positive results with Lumac and Bac-T-Screen were 20.6 and 22.3%, respectively. False-negative results for cultures with greater than or equal to 10(4) CFU/ml were 22.0% with AMS, 29.4% with Lumac, and 25.5% with Bac-T-Screen, and false-negative results for cultures with greater than or equal to 10(5) CFU/ml were 29.6% with AMS, 9.9% with Lumac, and 7.0% with Bac-T-Screen. For each system, greater than 70% of false-negatives at greater than or equal to 10(5) CFU/ml consisted of mixed or pure cultures of common contaminants. With any of these screening methods, a clinically significant isolate at greater than or equal to 10(5) CFU/ml would rarely be missed (less than or equal to 1.7% for all systems). A cost-effective and rapid approach to urine microbiology could consist of screening out negative specimens by either Lumac or Bac-T-Screen and processing only screen-positive specimens by the AMS.
Similar articles
-
Clinical evaluation of the Lumac bioluminescence method for screening urine specimens.J Clin Microbiol. 1985 Jul;22(1):19-22. doi: 10.1128/jcm.22.1.19-22.1985. J Clin Microbiol. 1985. PMID: 4019739 Free PMC article.
-
Evaluation of the MS-2 and Lumac systems for the rapid screening of urine specimens.Am J Clin Pathol. 1984 May;81(5):629-33. doi: 10.1093/ajcp/81.5.629. Am J Clin Pathol. 1984. PMID: 6372433
-
Use of rapid screening tests in processing urine specimens by conventional culture and the AutoMicrobic system.J Clin Microbiol. 1985 May;21(5):783-7. doi: 10.1128/jcm.21.5.783-787.1985. J Clin Microbiol. 1985. PMID: 3889047 Free PMC article.
-
Evaluation of a two-minute test for urine screening.J Clin Microbiol. 1983 Sep;18(3):697-701. doi: 10.1128/jcm.18.3.697-701.1983. J Clin Microbiol. 1983. PMID: 6355154 Free PMC article.
-
Automated methods for detection of bacteriuria.Am J Med. 1983 Jul 28;75(1B):71-8. doi: 10.1016/0002-9343(83)90075-x. Am J Med. 1983. PMID: 6192716 Review.
Cited by
-
Visual and clinical analysis of Bac-T-Screen urine screen results.J Clin Microbiol. 1988 Nov;26(11):2382-6. doi: 10.1128/jcm.26.11.2382-2386.1988. J Clin Microbiol. 1988. PMID: 3235667 Free PMC article.
-
Rapid screening for bacteriuria in pregnancy.Infection. 1992 Jan-Feb;20(1):9-11. doi: 10.1007/BF01704885. Infection. 1992. PMID: 1563815 Clinical Trial.
-
FiltraCheck-UTI, a rapid, disposable system for detection of bacteriuria.J Clin Microbiol. 1987 May;25(5):926-8. doi: 10.1128/jcm.25.5.926-928.1987. J Clin Microbiol. 1987. PMID: 3584428 Free PMC article.
-
Rapid bioluminescence method for bacteriuria screening.J Clin Microbiol. 1989 Apr;27(4):716-20. doi: 10.1128/jcm.27.4.716-720.1989. J Clin Microbiol. 1989. PMID: 2656743 Free PMC article.
-
Evaluation of BACTEC system for urine culture screening.J Clin Microbiol. 1989 Jan;27(1):118-9. doi: 10.1128/jcm.27.1.118-119.1989. J Clin Microbiol. 1989. PMID: 2643615 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources