Log file-based quality assurance method for respiratory gating system
- PMID: 40197704
- PMCID: PMC12256268
- DOI: 10.1002/acm2.70101
Log file-based quality assurance method for respiratory gating system
Abstract
Background: As medical linear accelerator technology advances, enabling higher dose rate deliveries, hypofractionation regimens has increased. This necessitates respiratory gating systems that synchronize radiation delivery with tumor position, requiring simple rigorous quality assurance (QA) to ensure treatment accuracy and patient safety.
Purpose: This study aimed to propose log-based QA for respiratory-gated radiation therapy using the respiratory gating system and treatment machine.
Methods: 4D CT scans were performed with a Varian motion phantom using a Varian Respiratory Gating for Scanner (RGSC). A treatment plan using 25%-75% respiratory phases with 100 MU was created and delivered to a solid water phantom. Treatment logs containing respiratory signals, beam on/off flags, and frame information were extracted from the treatment planning system's offline review. Log file analyses were conducted using in-house softwares to assess temporal synchronization between respiratory phases and beam triggers. Output measurements using a calibrated ion chamber (FC65G) were performed to evaluate dosimetric accuracy. Additionally, EPID images were acquired in cine mode and analyzed frame-by-frame to independently verify beam delivery timing.
Results: Log file analysis revealed precise temporal synchronization, with mean time differences of 0.03 s ± 0.05 s between the planned 25% phase and beam-on, and -0.04 s ± 0.05 s between 75% phase and beam-off. The log-derived beam-on duration (2.61 s ± 0.02 s) closely matched the planned duration (2.66 s ± 0.00 s). Three-month log data showed consistent temporal accuracy, with trigger-on times remaining stable at 2.60 s ± 0.01 s across all measurements. Supporting ion chamber measurements confirmed dosimetric agreement between gating and non-gating modes (difference: 0.05 cGy ± 0.09 cGy) CONCLUSIONS: The proposed log file-based QA method demonstrated high accuracy and reproducibility in assessing respiratory gating performance. This approach provides an efficient, objective method for standardizing QA procedures in respiratory-gated radiation therapy, enhancing treatment accuracy and patient safety.
Keywords: Log file‐based analysis; RGSC (Respiratory Gating for Scanner) system; quality assurance (QA); radiation therapy; respiratory gating.
© 2025 The Author(s). Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.
Conflict of interest statement
The authors declare no conflicts of interest.
Figures
References
-
- Choun HJ, Kim J‐i, Park JM, Son J. Development of a breath control training system for breath‐hold techniques and respiratory‐gated radiation therapy. Prog Med Phys. 2022;33(4):136‐141. doi: 10.14316/pmp.2022.33.4.136 - DOI
-
- Keall PJ, Mageras GS, Balter JM, et al. The management of respiratory motion in radiation oncology report of AAPM Task Group 76 a. Med Phys. 2006;33(10):3874‐3900. - PubMed
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical
Research Materials