Waveform Score Influences the Outcome Metrics of the Ocular Response Analyzer in Patients with Keratoconus and in Healthy Controls
- PMID: 40210210
- PMCID: PMC12187544
- DOI: 10.1080/02713683.2025.2489607
Waveform Score Influences the Outcome Metrics of the Ocular Response Analyzer in Patients with Keratoconus and in Healthy Controls
Abstract
Purpose: To test whether the intraocular-pressure (IOP) and biomechanical outcome metrics from the Ocular Response Analyzer (ORA) differ between the measurement with the highest waveform score and the average of four measurements of any waveform score in participants with keratoconus and in controls.
Methods: Patients with diagnosed keratoconus and healthy controls were recruited prospectively. Four measurements were made using a third-generation ORA. Goldmann-correlated IOP (IOPg), corneal-compensated IOP (IOPcc), corneal hysteresis (CH), corneal resistance factor (CRF), waveform score, and six waveform parameters (p1area, p2area, w1, w2, h1, and h2) were considered as outcome metrics. In the left eye, outcomes from the measurement with the highest waveform score were compared against averaged outcomes from four measurements of any waveform score using either paired t-tests or Wilcoxon signed-rank tests. Receiver operating characteristic (ROC) curves tested ability of both data-selection approaches to differentiate the cohorts.
Results: One-hundred ninety-two (N = 192) participants were enrolled. In the control cohort (n = 145), waveform score, CH, p2area, and h2 were all significantly greater for the best-waveform-score measurement than for the average-waveform-score measurement. IOPcc and w2 were significantly less for the best-waveform-score measurement than for the average-waveform-score measurement. In the keratoconus cohort (n = 47), waveform score, p1area, p2area, h1, and h2 were all significantly greater for the best-waveform-score measurement than for the average-waveform-score measurement. W2 was significantly less for the best-waveform-score measurement than for the average-waveform-score measurement. The area under the ROC curve was high for both data-selection approaches.
Conclusion: In general, the best measurement had higher and narrower waveform peaks than the averaged measurement, which suggests better alignment between the device and the eye in the former than in the latter. Thus, making multiple measurements and then analyzing the one with the single highest quality may be preferred to analyzing the average of the group.
Keywords: cornea; corneal hysteresis; intraocular pressure; ocular biomechanics; waveform parameters.
Conflict of interest statement
Declaration of Interest Statement
PTY is a consultant for CooperVision, Inc. MMF, MN, and AMM have no competing interests to declare. CJR is a consultant for Ziemer Ophthalmic Systems AG and for Oculus Optikerate GmbH. Reichert provided technical expertise for the study.
Similar articles
-
Intraocular pressure and ocular biomechanical parameters vary between generations of the ocular response analyzer in healthy and ectatic eyes.Front Med (Lausanne). 2025 Jul 24;12:1605641. doi: 10.3389/fmed.2025.1605641. eCollection 2025. Front Med (Lausanne). 2025. PMID: 40776931 Free PMC article.
-
ORA waveform-derived biomechanical parameters to distinguish normal from keratoconic eyes.Arq Bras Oftalmol. 2013 Mar-Apr;76(2):111-7. doi: 10.1590/s0004-27492013000200011. Arq Bras Oftalmol. 2013. PMID: 23828472
-
Assessment of Tonometry Methods in Keratoconic Eyes Following Intracorneal Ring Segments Implantation: A Comparative Study.Korean J Ophthalmol. 2025 Jun;39(3):231-240. doi: 10.3341/kjo.2025.0016. Epub 2025 May 7. Korean J Ophthalmol. 2025. PMID: 40329769 Free PMC article.
-
Systemic pharmacological treatments for chronic plaque psoriasis: a network meta-analysis.Cochrane Database Syst Rev. 2017 Dec 22;12(12):CD011535. doi: 10.1002/14651858.CD011535.pub2. Cochrane Database Syst Rev. 2017. Update in: Cochrane Database Syst Rev. 2020 Jan 9;1:CD011535. doi: 10.1002/14651858.CD011535.pub3. PMID: 29271481 Free PMC article. Updated.
-
Systemic pharmacological treatments for chronic plaque psoriasis: a network meta-analysis.Cochrane Database Syst Rev. 2021 Apr 19;4(4):CD011535. doi: 10.1002/14651858.CD011535.pub4. Cochrane Database Syst Rev. 2021. Update in: Cochrane Database Syst Rev. 2022 May 23;5:CD011535. doi: 10.1002/14651858.CD011535.pub5. PMID: 33871055 Free PMC article. Updated.
Cited by
-
Intraocular pressure and ocular biomechanical parameters vary between generations of the ocular response analyzer in healthy and ectatic eyes.Front Med (Lausanne). 2025 Jul 24;12:1605641. doi: 10.3389/fmed.2025.1605641. eCollection 2025. Front Med (Lausanne). 2025. PMID: 40776931 Free PMC article.
References
-
- Ehrlich JR, Haseltine S, Shimmyo M, Radcliffe NM. Evaluation of agreement between intraocular pressure measurements using goldmann applanation tonometry and goldmann correlated intraocular pressure by reichert’s ocular response analyser. Eye (Lond). 2010;24(10):1555–1560. - PubMed
-
- Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg. 2005;31(1):156–162. - PubMed
-
- Medeiros FA, Weinreb RN. Evaluation of the influence of corneal biomechanical properties on intraocular pressure measurements using the ocular response analyzer. J Glaucoma. 2006;15(5):364–370. - PubMed
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources