Improving tissue contrast visualization in two-point Dixon MRI using dark-fat processing: application in clinical knee imaging
- PMID: 40220143
- DOI: 10.1007/s00256-025-04925-2
Improving tissue contrast visualization in two-point Dixon MRI using dark-fat processing: application in clinical knee imaging
Abstract
Objective: To improve tissue contrast visualization in water-only images from two-point turbo spin-echo (TSE) Dixon MRI using dark-fat image processing and evaluate in retrospectively acquired clinical knee images.
Materials and methods: Clinical knee MRI datasets from 36 patients were retrospectively compiled under IRB approval. The dark-fat water-only images were generated and compared with the conventional water-only images from two-point TSE-Dixon MRI. The water-only images from 20 patients were analyzed: (i) qualitatively by two experienced musculoskeletal radiologists, for overall image quality, fat suppression, anatomical visualization, and the absence of artifacts, using Wilcoxon rank tests, and (ii) quantitatively using apparent signal-to-noise ratio (aSNR) and apparent contrast-to-noise ratio (aCNR) from regions of interest (ROIs) with paired t-test. In the remaining 16 patients, the number of cruciate ligament tears identified on MRI was tabulated and compared with arthroscopy results.
Results: Compared to conventional water-only images, dark-fat water-only images showed improved fat suppression (reader 1, P < 0.0001; reader 2, P < 0.05), improved overall image quality by reader 1 (P < 0.001), and improved visualization of the sciatic nerve for reader 1 (P < 0.01), while there were no significant differences for the other anatomical structures. aSNR was significantly reduced for bone marrow fat (P < 0.0001), and aCNR between meniscus and articular cartilage was significantly improved (P < 0.0001). There were a higher number of posterior cruciate ligament tears identified in MRI compared to arthroscopy.
Conclusion: Dark-fat processed water-only images improved fat suppression and tissue contrast visualization, particularly sciatic nerves, compared to conventional water-only images from two-point TSE-Dixon MR imaging.
Keywords: Arthroscopy; Dark-fat; Dixon; Knee imaging; TSE-Dixon.
© 2025. The Author(s), under exclusive licence to International Skeletal Society (ISS).
Conflict of interest statement
Declarations. Ethics approval and consent to participate: This study was performed in accordance with the institutional review board (IRB) protocols. The retrospective IRB protocol has waived informed consent from the subjects. Conflict of interest: The authors declare no competing interests.
Similar articles
-
Free-Breathing High-Resolution, Swap-Free, and Motion-Corrected Water/Fat Separation in Pediatric Abdominal MRI.Invest Radiol. 2024 Dec 1;59(12):805-812. doi: 10.1097/RLI.0000000000001092. Epub 2024 Jun 10. Invest Radiol. 2024. PMID: 38857418
-
Improved Brain Tumor Conspicuity at 3 T Using Dark Blood, Fat-Suppressed, Dixon Unbalanced T1 Relaxation-Enhanced Steady-State MRI.Invest Radiol. 2023 Sep 1;58(9):641-648. doi: 10.1097/RLI.0000000000000964. Invest Radiol. 2023. PMID: 36822675 Free PMC article.
-
Accelerated High-Resolution Deep Learning Reconstruction Turbo Spin Echo MRI of the Knee at 7 T.Invest Radiol. 2024 Dec 1;59(12):831-837. doi: 10.1097/RLI.0000000000001095. Epub 2024 Jul 4. Invest Radiol. 2024. PMID: 38960863
-
Platelet-rich therapies for musculoskeletal soft tissue injuries.Cochrane Database Syst Rev. 2013 Dec 23;(12):CD010071. doi: 10.1002/14651858.CD010071.pub2. Cochrane Database Syst Rev. 2013. Update in: Cochrane Database Syst Rev. 2014 Apr 29;(4):CD010071. doi: 10.1002/14651858.CD010071.pub3. PMID: 24363098 Updated.
-
Retention versus sacrifice of the posterior cruciate ligament in total knee arthroplasty for treating osteoarthritis.Cochrane Database Syst Rev. 2013 Oct 11;2013(10):CD004803. doi: 10.1002/14651858.CD004803.pub3. Cochrane Database Syst Rev. 2013. PMID: 24114343 Free PMC article.
References
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical