The FreeStyle Libre 2 system demonstrates limited analytical accuracy without resulting in detrimental clinical decisions
- PMID: 40280161
- DOI: 10.2460/ajvr.25.02.0055
The FreeStyle Libre 2 system demonstrates limited analytical accuracy without resulting in detrimental clinical decisions
Abstract
Objective: To evaluate the performance of a second-generation flash glucose monitoring system (FGMS; FreeStyle Libre 2) in nondiabetic dogs during rapidly induced hypoglycemia.
Methods: In a prospective, observational study, regular insulin was administered to 23 healthy dogs to induce hypoglycemia, with subsequent correction. Serial measurements of interstitial glucose with FGMS and blood glucose (BG) with a portable BG monitor (PBGM) were recorded every 10 minutes for 90 minutes. At each time point, blood was also collected for gold standard BG measurement via clinical chemistry analyzer. Analytical and clinical accuracy was determined by fulfillment of International Organization for Standardization 15197:2013 criteria.
Results: For BG values < 100 mg/dL, 67.2% of PBGM measurements and 50.9% of FGMS measurements were within ± 15 mg/dL of the reference method. In Parkes error grid analysis, 94.9% of PBGM and 88.6% of FGMS measurements fell within zones A and B. Readings within higher-risk zones were rare, with no PBGM values and 0.6% of FGMS values in zone D and neither producing readings in zone E.
Conclusions: In this model of rapidly induced hypoglycemia, the FGMS demonstrated limited analytical accuracy compared to the reference standard. However, it rarely produced glucose readings likely to lead to detrimental clinical decisions.
Clinical relevance: The FreeStyle Libre 2 is clinically useful for monitoring glucose trends in dogs, though results should be interpreted with caution during periods of rapid glycemic change.
Keywords: FreeStyle Libre 2; canine; diabetes mellitus; flash glucose monitoring system; interstitial glucose.
Similar articles
-
Evaluation of the Accuracy of FreeStyle Libre 2 for Glucose Monitoring in White New Zealand Rabbits.Vet Med Sci. 2025 Jan;11(1):e70166. doi: 10.1002/vms3.70166. Vet Med Sci. 2025. PMID: 39688518 Free PMC article.
-
Evaluation of a flash glucose monitoring system in nondiabetic dogs with rapidly changing blood glucose concentrations.J Vet Intern Med. 2021 Nov;35(6):2628-2635. doi: 10.1111/jvim.16273. Epub 2021 Oct 2. J Vet Intern Med. 2021. PMID: 34599607 Free PMC article.
-
Two human portable glucometers and a veterinary point-of-care glucometer correlate well with a reference laboratory chemistry analyzer for measurement of blood glucose concentrations in dogs.Am J Vet Res. 2025 Apr 2;86(6):ajvr.24.10.0317. doi: 10.2460/ajvr.24.10.0317. Print 2025 Jun 1. Am J Vet Res. 2025. PMID: 40179951
-
Continuous glucose monitoring systems for type 1 diabetes mellitus.Cochrane Database Syst Rev. 2012 Jan 18;1(1):CD008101. doi: 10.1002/14651858.CD008101.pub2. Cochrane Database Syst Rev. 2012. PMID: 22258980 Free PMC article.
-
Integrated sensor-augmented pump therapy systems [the MiniMed® Paradigm™ Veo system and the Vibe™ and G4® PLATINUM CGM (continuous glucose monitoring) system] for managing blood glucose levels in type 1 diabetes: a systematic review and economic evaluation.Health Technol Assess. 2016 Feb;20(17):v-xxxi, 1-251. doi: 10.3310/hta20170. Health Technol Assess. 2016. PMID: 26933827 Free PMC article.
Cited by
-
Glycated Albumin and Continuous Glucose Monitoring Metrics in Dogs with Diabetes Mellitus: A Pilot Study.Animals (Basel). 2025 Jul 8;15(14):2004. doi: 10.3390/ani15142004. Animals (Basel). 2025. PMID: 40723467 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical