Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2025 Jul 1;36(5):191-198.
doi: 10.1097/MBC.0000000000001363. Epub 2025 Apr 23.

Evaluation of a 4 min 4000 g centrifugation protocol for routine coagulation assays

Affiliations

Evaluation of a 4 min 4000 g centrifugation protocol for routine coagulation assays

Sylvain Lamoine et al. Blood Coagul Fibrinolysis. .

Abstract

Centrifugation is a critical step in sample preparation and accounts for an important part of turnaround time. This step is further critical for hemostasis, which requires a low platelet count to produce reliable results. For automated laboratories, centrifugation can represent a bottleneck and thus a shorter centrifugation time would benefit tube flow and turnaround time. We compared a rapid centrifugation protocol (4000 g 4 min) with the recommended protocol (2200 g 15 min) at two different centers, after one or two centrifugation cycles. The effect of each protocol was assessed on the platelet count at every step to verify the capacity of the protocol to yield platelet-poor plasma (PPP). Results on 16 coagulation parameters were compared to verify the reliability of rapid centrifugation. In one center, a consecutive two-cycle centrifugation had been tested on platelet count. A single centrifugation cycle, using the rapid protocol, produced plasma with increased residual platelets compared with the Groupe Etude sur l'Hémostase et la Thrombose (GEHT) protocol. Despite this difference, the coagulation results were interchangeable between the protocols. In addition, a second centrifugation cycle produces plasma with a mean residual platelet less than or equal to 10 × 10 9 /l. A single cycle of rapid centrifugation can be used to assess prothrombin time (PT), activated partial thromboplastin time (aPTT), aPTT kaolin, thrombin time (TT), fibrinogen, antithrombin (AT), D-dimers, anti-Xa, factor II (FII), factor V (FV), factor VII (FVII), and factor X (FX). For frozen plasmas, a double-cycle followed by a third cycle should be performed to ensure that 100% of samples contain less than 10 × 10 9 /l platelets.

Keywords: coagulation factors; hemostasis; platelet count; rapid centrifugation; turnaround time.

PubMed Disclaimer

References

    1. Kitchen S, Adcock DM, Dauer R, Kristoffersen AH, Lippi G, Mackie I, et al. International Council for Standardization in Haematology (ICSH) recommendations for processing of blood samples for coagulation testing. Int J Lab Hematol 2021; 43:1272–1283.
    1. Nelson S, Pritt A, Marlar RA. Rapid preparation of plasma for “Stat” coagulation testing. Arch Pathol Lab Med 1994; 118:175–176.
    1. Söderström AC, Nybo M, Nielsen C, Vinholt PJ. The effect of centrifugation speed and time on preanalytical platelet activation. Clin Chem Lab Med 2016; 54:1913–1920.
    1. Suchsland J, Friedrich N, Grotevendt A, Kallner A, Lüdemann J, Nauck M, Petersmann A. Optimizing centrifugation of coagulation samples in laboratory automation. Clin Chem Lab Med CCLM 2014; 52:1187–1191.
    1. Boissier E, Sévin-Allouet M, Le Thuaut A, De Gaalon S, Trossaërt M, Rozec B, et al. A 2-min at 4500 g rather than a 15-min at 2200 g centrifugation does not impact the reliability of 10 critical coagulation assays. Clin Chem Lab Med CCLM 2017; 55:e118–e121.

Publication types

LinkOut - more resources