Real-time tracking of lung tumors using a 1.5T Elekta Unity MR-Linac: first clinical experiences
- PMID: 40313246
- PMCID: PMC12044338
- DOI: 10.3389/fonc.2025.1569428
Real-time tracking of lung tumors using a 1.5T Elekta Unity MR-Linac: first clinical experiences
Abstract
There is a growing interest in the application of MRI-guided adaptive radiotherapy (MRIgART) to improve the treatment of lung tumors. Motion management plays a central role to better localize these tumors and minimize toxicities to surrounding healthy tissue. Elekta recently released their comprehensive motion management system for the Unity 1.5T MR-Linac, enabling real-time tracking and predictive gating using MR cine imaging during treatment. We report our first clinical experiences using this gating technology to treat patients with central lung tumors adjacent to critical structures on the Unity MR-Linac using both conventional and SBRT fractionations. A surrogate structure was tracked if the primary target could not be visualized. Beam gating was automatically performed if the system detected the target moving outside of the defined gating envelope, which was set as the planning target volume. Integrating this technology has profoundly impacted our MRIgART program; we have been able to treat challenging lung tumor patients with greater precision and have begun to reduce our PTV margins, leading to improved dose sparing of nearby critical structures.
Keywords: MR-Linac; adaptive; gating; lung; motion management.
Copyright © 2025 Smith, Allen, Rusu, St-Aubin and Hyer.
Conflict of interest statement
JS-A reports honorarium and research funding from Elekta unrelated to this work. DH discloses a consulting relationship with Elekta and research funding from Elekta unrelated to this work. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Figures


Similar articles
-
Surrogate gating strategies for the Elekta Unity MR-Linac gating system.J Appl Clin Med Phys. 2025 Feb;26(2):e14566. doi: 10.1002/acm2.14566. Epub 2024 Nov 14. J Appl Clin Med Phys. 2025. PMID: 39540669 Free PMC article.
-
Commissioning of a motion management system for a 1.5T Elekta Unity MR-Linac: A single institution experience.J Appl Clin Med Phys. 2025 Apr;26(4):e70005. doi: 10.1002/acm2.70005. Epub 2025 Feb 16. J Appl Clin Med Phys. 2025. PMID: 39955657 Free PMC article.
-
Experimental demonstration of real-time cardiac physiology-based radiotherapy gating for improved cardiac radioablation on an MR-linac.Med Phys. 2024 Apr;51(4):2354-2366. doi: 10.1002/mp.17024. Epub 2024 Mar 13. Med Phys. 2024. PMID: 38477841
-
Stereotactic Magnetic Resonance-Guided Adaptive and Non-Adaptive Radiotherapy on Combination MR-Linear Accelerators: Current Practice and Future Directions.Cancers (Basel). 2023 Mar 30;15(7):2081. doi: 10.3390/cancers15072081. Cancers (Basel). 2023. PMID: 37046741 Free PMC article. Review.
-
SBRT targets that move with respiration.Phys Med. 2018 Dec;56:19-24. doi: 10.1016/j.ejmp.2018.10.021. Epub 2018 Nov 17. Phys Med. 2018. PMID: 30527085 Review.
References
-
- Raaymakers BW, Jurgenliemk-Schulz IM, Bol GH, Glitzner M, Kotte A, van Asselen B, et al. . First patients treated with a 1.5 T MRI-Linac: clinical proof of concept of a high-precision, high-field MRI guided radiotherapy treatment. Phys Med Biol. (2017) 62:L41–50. doi: 10.1088/1361-6560/aa9517 - DOI - PubMed