Pulse oximetry in pediatric intensive care: comparison with measured saturations and transcutaneous oxygen tension
- PMID: 4032131
- DOI: 10.1016/s0022-3476(85)80507-2
Pulse oximetry in pediatric intensive care: comparison with measured saturations and transcutaneous oxygen tension
Abstract
We evaluated a new pulse oximeter designed to monitor beat-to-beat arterial oxygen saturation (SaO2) and compared the monitored SaO2 with arterial samples measured by co-oximetry. In 40 critically ill children (112 data sets) with a mean age of 3.9 years (range 1 day to 19 years), SaO2 ranged from 57% to 100%, and PaO2 from 27 to 128 mm Hg, heart rates from 85 to 210 beats per minute, hematocrit from 20% to 67%, and fetal hemoglobin levels from 1.3% to 60%; peripheral temperatures varied between 26.5 degrees and 36.5 degrees C. Linear correlation analysis revealed a good agreement between simultaneous pulse oximeter values and both directly measured SaO2 (r = 0.95) and that calculated from measured arterial PaO2 (r = 0.95). The device detected several otherwise unrecognized drops in SaO2 but failed to function in four patients with poor peripheral perfusion secondary to low cardiac output. Simultaneous measurements with a tcPO2 electrode showed a similarly good correlation with PaO22 (r = 0.91), but the differences between the two measurements were much wider (mean 7.1 +/- 10.3 mm Hg, range -14 to +49 mm Hg) than the differences between pulse oximeter SaO2 and measured SaO2 (1.5% +/- 3.5%, range -7.5% to -9%) and were not predictable. We conclude that pulse oximetry is a reliable and accurate noninvasive device for measuring saturation, which because of its rapid response time may be an important advance in monitoring changes in oxygenation and guiding oxygen therapy.
Similar articles
-
Pulse oximetry and transcutaneous oxygen tension for detection of hypoxemia in critically ill infants and children.Adv Exp Med Biol. 1987;220:159-64. doi: 10.1007/978-1-4613-1927-6_28. Adv Exp Med Biol. 1987. PMID: 3673758
-
Pulse oximetry for continuous oxygen monitoring in sick newborn infants.J Pediatr. 1986 Dec;109(6):1052-6. doi: 10.1016/s0022-3476(86)80298-0. J Pediatr. 1986. PMID: 2431130
-
Pulse oximetry in very low birth weight infants with acute and chronic lung disease.Pediatrics. 1987 Apr;79(4):612-7. Pediatrics. 1987. PMID: 2434913
-
Pulse oximetry for hypoxemia: a warning to users and manufacturers.Intensive Care Med. 1989;15(8):540-2. doi: 10.1007/BF00273571. Intensive Care Med. 1989. PMID: 2691541 Review.
-
Noninvasive monitoring of oxygenation in infants and children: practical considerations and areas of concern.Pediatrics. 1994 May;93(5):737-46. Pediatrics. 1994. PMID: 8165071 Review.
Cited by
-
Oxygen saturation and breathing patterns in infancy. 2: Preterm infants at discharge from special care.Arch Dis Child. 1991 May;66(5):574-8. doi: 10.1136/adc.66.5.574. Arch Dis Child. 1991. PMID: 2039244 Free PMC article.
-
Simple and noninvasive indicator of pulmonary gas exchange impairment using pulse oximetry.J Clin Monit. 1989 Apr;5(2):82-6. doi: 10.1007/BF01617878. J Clin Monit. 1989. PMID: 2656926
-
Transcutaneous oxygen saturation in sleeping infants: prone and supine.Arch Dis Child. 1990 May;65(5):524-6. doi: 10.1136/adc.65.5.524. Arch Dis Child. 1990. PMID: 2357093 Free PMC article. Clinical Trial.
-
Continuous neonatal evaluation in the delivery room by pulse oximetry.J Clin Monit. 1987 Apr;3(2):96-100. doi: 10.1007/BF00858357. J Clin Monit. 1987. PMID: 3585439
-
Continuous measurement of cardiac output by the Fick principle in infants and children: comparison with the thermodilution method.Intensive Care Med. 1996 May;22(5):467-71. doi: 10.1007/BF01712169. Intensive Care Med. 1996. PMID: 8796404 Review.
Publication types