Comparative Study
. 1985 Aug;20(8):536-41.
doi: 10.1007/BF02534895.

Quantitating heart lipids: comparison of results obtained using the Iatroscan method with those from phosphorus and gas chromatographic techniques

Comparative Study

Quantitating heart lipids: comparison of results obtained using the Iatroscan method with those from phosphorus and gas chromatographic techniques

J K Kramer et al. Lipids. 1985 Aug.

Abstract

The precision and accuracy of the Iatroscan method was evaluated by comparing the results obtained with established phosphorus and gas chromatographic techniques. A complete lipid class analysis of rat heart lipids was chosen in order to evaluate the performance of the Iatroscan method for biological samples which contained both neutral lipids and phospholipids. A partial scan and repeat development with chloroform/methanol/water (68.5:29:2.5) was introduced to achieve consistently good separations of the phospholipids on the Chromarods in the Iatroscan method. The results showed that the precision of the Iatroscan method for some lipid classes was comparable to that of phosphorus or gas chromatographic techniques, while for other lipid classes it was lower. Compared to the data obtained using the phosphorus method, the Iatroscan data were generally similar, while the gas chromatographic method generally gave lower values. These findings, together with the advantages of time required for analysis, size of sample, and universality of detection, suggest that the Iatroscan is a valuable complementary method for complex lipid analyses.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Chromatogr. 1979 Oct 11;164(2):222-7 - PubMed
    1. J Chromatogr. 1980 May 30;193(3):504-10 - PubMed
    1. J Nutr. 1983 Dec;113(12):2442-54 - PubMed
    1. Lipids. 1980 Oct;15(10):872-5 - PubMed
    1. Clin Chim Acta. 1979 Apr 16;93(2):173-80 - PubMed

Publication types

LinkOut - more resources