How Patients' Lifestyle Affects the Wear of Hip Implants: An In-Silico Study
- PMID: 40372250
- DOI: 10.1002/jor.26098
How Patients' Lifestyle Affects the Wear of Hip Implants: An In-Silico Study
Abstract
The validation of new hip replacement designs traditionally relies on lengthy and costly In-Vitro wear tests that replicate basic In-Vivo conditions, as the simplified walking recommended in the ISO 14242 guidelines. These tests overlook the diverse motor tasks and lifestyle differences among patients. This study seeks to establish the foundation for In-Silico clinical trials of total hip replacements, enabling wear simulations of patients with different lifestyles, not feasible with In-Vitro tests. The impact of diverse kinematic and loading histories on the wear of metal-on-plastic hip replacements is investigated in a novel way, considering the combined effect of six daily activities (e.g., walking, fast walking, sit/stand, stairs up/down, lunging), different activity frequencies across five patient profiles (from sedentary elderly to active young), and the effect of load sequence. The results reveal that both the type and frequency of motor tasks significantly influence implant wear. The most critical tasks and at-risk patients were stair climbing and the most active individuals, regardless of age. Load sequence also plays a key role in long-term wear predictions. Accuracy and computational cost were balanced by simulating walking, stair climbing, and sit/stand cycles, ensuring equivalent wear to a complete motor task sequence. ISO standards conditions notably tend to underestimate volumetric wear by up to 60% compared to the simulated patient types. They also fail to predict realistic wear patterns for activities like squatting and lunging where edge contact occurs.
Keywords: FEM wear modeling; daily activity; loading history; metal‐on‐plastic; patient lifestyle; wear maps.
© 2025 Orthopaedic Research Society.
Similar articles
-
Effects of Physiological Loading from Patient-Derived Activities of Daily Living on the Wear of Metal-on-Polymer Total Hip Replacements.Bioengineering (Basel). 2025 Jun 16;12(6):663. doi: 10.3390/bioengineering12060663. Bioengineering (Basel). 2025. PMID: 40564479 Free PMC article.
-
Hip contact forces can be directed outside of a well-oriented cup during common activities; implications for implant testing.Clin Biomech (Bristol). 2025 Jul;127:106576. doi: 10.1016/j.clinbiomech.2025.106576. Epub 2025 Jun 4. Clin Biomech (Bristol). 2025. PMID: 40489939
-
Corticosteroids for the treatment of Duchenne muscular dystrophy.Cochrane Database Syst Rev. 2016 May 5;2016(5):CD003725. doi: 10.1002/14651858.CD003725.pub4. Cochrane Database Syst Rev. 2016. PMID: 27149418 Free PMC article.
-
Total hip replacement and surface replacement for the treatment of pain and disability resulting from end-stage arthritis of the hip (review of technology appraisal guidance 2 and 44): systematic review and economic evaluation.Health Technol Assess. 2015 Jan;19(10):1-668, vii-viii. doi: 10.3310/hta19100. Health Technol Assess. 2015. PMID: 25634033 Free PMC article.
-
Associations Between Skeletal Alignment and Biomechanical Symmetry Before and After Transfemoral Bone-anchored Limb Implantation.Clin Orthop Relat Res. 2025 May 1;483(5):902-914. doi: 10.1097/CORR.0000000000003344. Epub 2024 Dec 24. Clin Orthop Relat Res. 2025. PMID: 39745233
References
-
- J. T. Evans, J. P. Evans, R. W. Walker, A. W. Blom, M. R. Whitehouse, and A. Sayers, “How Long Does a Hip Replacement Last? A Systematic Review and Meta‐Analysis of Case Series and National Registry Reports With More Than 15 Years of Follow‐Up,” Lancet 393, no. 10172 (2019): 647–654, https://doi.org/10.1016/S0140-6736(18)31665-9.
-
- L. Wang, G. Isaac, R. Wilcox, A. Jones, and J. Thompson, “Finite Element Analysis of Polyethylene Wear in Total Hip Replacement: A Literature Review,” Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine 233, no. 11 (2019): 1067–1088, https://doi.org/10.1177/0954411919872630.
-
- BS ISO 14242‐1. Implants for Surgery, Wear of Total Hip‐Joint Prostheses. Part 1: Loading and Displacement Parameters for Wear‐Testing Machines and Corresponding Environmental Conditions for Test. Published Online 2014.
-
- M. Hadley, C. Hardaker, G. Isaac, and J. Fisher, “Wear of Different Materials for Total Hip Replacement Under Adverse Stop‐Dwell‐Start In Vitro Wear Simulation Conditions,” Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine 232, no. 12 (2018): 1261–1270, https://doi.org/10.1177/0954411918813385.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical