Validation of a Doppler technique for beat-to-beat measurement of cardiac output
- PMID: 4042540
- DOI: 10.1042/cs0690377
Validation of a Doppler technique for beat-to-beat measurement of cardiac output
Abstract
We have measured aortic flow in the ascending aorta in man with a Bach-Simpson BVM 202 blood velocity meter, and aortic root diameter by M-mode echocardiography, and thus derived beat-to-beat cardiac output (Q). We tested the technique in 21 patients (53 comparisons) with various cardiovascular problems against a thermodilution method, and in four normal subjects at rest and two levels of exercise (50 and 100 W) against a nitrous oxide rebreathing method. We obtained excellent overall correlation in a range of 0.5-10 litres/min (r = 0.98, n = 77, sy,x = 0.48 litre/min), the formula for the least squares regression being: (Q Doppler) = 0.95 (Q Thermodilution/N2O) + 0.11 litre/min. The Doppler signal is sufficiently noise-free to obtain maximum acceleration of flow from the first derivative of velocity.
Publication types
MeSH terms
Substances