Comparison of screw and plate osteosynthesis in advancement genioplasty: a finite element analysis study
- PMID: 40453407
- PMCID: PMC12126161
- DOI: 10.26650/eor.20241371296
Comparison of screw and plate osteosynthesis in advancement genioplasty: a finite element analysis study
Abstract
Purpose: The purpose of this study was to evaluate the distribution of stresses in screw and plate fixation systems during simulated advancement genioplasty using finite element analysis.
Materials and methods: A cone-beam computed tomography image of a patient was used to create three-dimensional virtual models of mandibular bone. Chin advancement of 8 mm was simulated following a horizontal osteotomy of the chin in a computer-aided design program. The distal segment was stabilized with two titanium mini-screws placed bilaterally in the first model and a single 4-hole titanium pre-bent chin plate placed centrally in the second model. The plate was fixed with four mini-screws, two in the proximal and two in the distal segment. All fixative appliances were submitted to 15 N force applied backwards to the lingual surface of the chin parallel to the occlusal plane and 7 N force applied upwards to the buccal surface of the chin perpendicular to the occlusal plane. The distributions of von Mises stresses and deformations in bone and titanium materials were evaluated.
Results: In the screw fixation system (22.52 MPa) higher stress values were observed compared to the plate fixation system (13.71 MPa). The deformation value was higher for the screw fixation system (0.021 mm) than the plate fixation system (0.0007 mm).
Conclusion: In advancement genioplasty, fixation with a single pre-bent centrally placed chin plate showed slightly better stabilization than fixation with two bilaterally placed bicortical screws. The stress values were within the physical strength limits of bone and titanium for both systems.
Keywords: stress distribution; Genioplasty; bone plate; bone screw; finite element analysis.
Copyright © 2025 European Oral Research.
Conflict of interest statement
Conflict of interest:The authors declared that they have no conflict of interest.
Figures









Similar articles
-
Force and deformation stresses in customized and non-customized plates during simulation of advancement genioplasty.J Craniomaxillofac Surg. 2017 Nov;45(11):1820-1827. doi: 10.1016/j.jcms.2017.08.021. Epub 2017 Aug 31. J Craniomaxillofac Surg. 2017. PMID: 28935483
-
Stability of the chin after advancement genioplasty using absorbable plate and screws with template devices.J Craniomaxillofac Surg. 2019 Oct;47(10):1498-1503. doi: 10.1016/j.jcms.2019.07.027. Epub 2019 Jul 29. J Craniomaxillofac Surg. 2019. PMID: 31402207
-
Evaluation of two stable internal fixation techniques in mandibular ramus sagittal osteotomy: a three-dimensional finite element study.Oral Maxillofac Surg. 2025 Feb 27;29(1):59. doi: 10.1007/s10006-025-01357-z. Oral Maxillofac Surg. 2025. PMID: 40011258
-
Fixation Methods for Mandibular Advancement and Their Effects on Temporomandibular Joint: A Finite Element Analysis Study.Biomed Res Int. 2020 Feb 21;2020:2810763. doi: 10.1155/2020/2810763. eCollection 2020. Biomed Res Int. 2020. PMID: 32185199 Free PMC article.
-
Stability of locking and conventional 2.0-mm miniplate/screw systems after sagittal split ramus osteotomy: finite element analysis.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009 Aug;108(2):174-7. doi: 10.1016/j.tripleo.2009.03.051. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009. PMID: 19615655
References
-
- Mortazavi H, Tabrizi R, Mohajerani H, Ozkan T. Evaluation of stability in advancement genioplasty: An 18-month follow-up. Shiraz Univ Dent J. 2010;10:61–4.
-
- Hofer O. Operation der prognathie and mikrogenie. Dtsch Zahn-, Mund-, und Kieferheilkunde 1942;9:121–32.
-
- Trauner R, Obwegeser H. The surgical correction of mandibular prognathism and retrognathia with consideration of genioplasty. I. Surgical procedures to correct mandibular prognathism and reshaping of the chin. Oral Surg Oral Med Oral Pathol. 1957. Jul;10(7):677–89. 10.1016/S0030-4220(57)80063-2 - DOI - PubMed
LinkOut - more resources
Full Text Sources