Optimizing the primary stability of dental implants in type IV bone: in-vitro comparison of machine-driven and ratcheting insertion protocols
- PMID: 40453411
- PMCID: PMC12126156
- DOI: 10.26650/eor.20241296069
Optimizing the primary stability of dental implants in type IV bone: in-vitro comparison of machine-driven and ratcheting insertion protocols
Abstract
Purpose: The objective of this study was to assess the effects of various implant insertion techniques on the primary stability of dental implants in both type II and type IV cadaveric bovine.
Materials and methods: A total of 48 dental implants (BEGO Semados RSX, BEGO Implant Systems GmbH and Co. KG, Germany) with a diameter of 3.75 mm and a length of 12 mm were used in the experiments. Bovine bone ribs were adjusted to mimic type II and type IV bone characteristics. Following the preparation of recipient sites, implants were inserted using three different protocols: machine-driven insertion (Standard group, Std group), ratchet insertion (Ratcheted, R Group), and a combination of both (Std + R group). The Osstell® Beacon device was used to record the implant stability quotient (ISQ) of each implant immediately after insertion. Two-way analysis of variance and Bonferroni tests were used for statistical evaluation.
Results: Bone type significantly influenced the ISQ values (p<0.05). However, when comparing insertion protocols separately for type II and type IV bone, no significant differences were observed. In type IV bone, both the Std group and R group exhibited significantly lower ISQ values compared to the same groups in type II bone (p<0.05 for each). Nevertheless, there were no significant differences in the ISQ values when employing the Std+R technique between the two types of bone.
Conclusion: Combining machine-driven and ratchet insertion techniques may prove beneficial in optimizing ISQ values in bovine samples simulating type IV bone.
Keywords: Dental implants; oral surgical procedures; osseointegration; resonance frequency analysis.
Copyright © 2025 European Oral Research.
Conflict of interest statement
Conflict of interest:The authors declared that they have no conflict of interest.
Figures
Similar articles
-
The Influence of the Degree of Dental Implant Insertion Compression on Primary Stability Measured by Resonance Frequency and Progressive Insertion Torque: In Vitro Study.Biomedicines. 2024 Dec 18;12(12):2878. doi: 10.3390/biomedicines12122878. Biomedicines. 2024. PMID: 39767784 Free PMC article.
-
Does Length, Diameter, or Bone Quality Affect Primary and Secondary Stability in Self-Tapping Dental Implants?J Oral Maxillofac Surg. 2016 Jul;74(7):1344-53. doi: 10.1016/j.joms.2016.03.011. Epub 2016 Mar 22. J Oral Maxillofac Surg. 2016. PMID: 27070843
-
In vitro comparison of resonance frequency analysis devices to evaluate implant stability of narrow diameter implants at varying drilling speeds in dense artificial bone blocks.Clin Implant Dent Relat Res. 2019 Oct;21(5):1023-1027. doi: 10.1111/cid.12842. Epub 2019 Aug 29. Clin Implant Dent Relat Res. 2019. PMID: 31464362
-
Effects of Drilling Protocol and Bone Density on the Stability of Implants According to Different Macrogeometries of the Implant Used: Results of an In Vitro Study.Int J Oral Maxillofac Implants. 2020 Sep/Oct;35(5):955-964. doi: 10.11607/jomi.8176. Int J Oral Maxillofac Implants. 2020. PMID: 32991646
-
A comparison of bone bed preparation with laser and conventional drill on the relationship between implant stability quotient (ISQ) values and implant insertion variables.J Adv Prosthodont. 2010 Dec;2(4):148-53. doi: 10.4047/jap.2010.2.4.148. Epub 2010 Dec 31. J Adv Prosthodont. 2010. PMID: 21264194 Free PMC article.
References
LinkOut - more resources
Full Text Sources
Research Materials