Systematic Review of the Impact of Magnetic Resonance Imaging on Diametric Magnet Cochlear Implants
- PMID: 40467098
- DOI: 10.1097/MAO.0000000000004556
Systematic Review of the Impact of Magnetic Resonance Imaging on Diametric Magnet Cochlear Implants
Abstract
Objective: Review the properties, safety, and artifacts of current cochlear implants (CI) with freely rotatable magnets during MRI scans.
Databases reviewed: PubMed, Cochrane, and Web of Science; last accessed September 2023.
Methods: The study was BSMH IRB exempted. MESH term "magnetic resonance imaging" and phrases of "Med-el Synchrony," "Cochlear Nucleus Profile Plus," "Advanced Bionics HiRes Ultra 3D," "rotatable magnet cochlear," "rotatable magnet," "diametric magnet," "diametric magnet cochlear implant," and "MRI-safe cochlear implant" using "AND" function. Non-English, non-human, non-implanted CI, and cadaveric studies were excluded. Methodological quality was assessed using the Mixed Methods Appraisal Tool, and quality of the evidence was evaluated using the GRADE criteria.
Results: The review included 18 studies. Fifteen assessed pain as a complication, with two reporting its occurrence: two cases for Synchrony and one for Ultra 3D. Demagnetization did not occur in the 12 studies that reported on it. One magnet displacement was noted with the Ultra 3D implant out of the 16 reporting studies. Artifacts were reported by 11 studies, with most scans retaining most to limited diagnostic usability. One Synchrony device required explantation due to artifact, and one was explanted before MRI.
Conclusion: Of the CI models reviewed with diametric magnets, Med-El Synchrony may enable MRI usage with fewer safety concerns when following the manufacturer's recommendations. However, clinical reports of freely rotatable CI safety during MRI scans are sparse and contain incomplete data. MRI artifacts from CI still pose concerns for image quality but can be minimized with appropriate implant location and scanning protocols.
Keywords: Cochlear implants; Diametric magnet; Magnetic resonance imaging; Patient safety; Self-aligning magnet.
Copyright © 2025, Otology & Neurotology, Inc.
Conflict of interest statement
The authors disclose no conflicts of interest.
Similar articles
-
Magnetic Resonance Imaging in Patients with Cochlear Implants without Magnet Removal: A Radiology-Administered Protocol to Enhance Operational Efficiency and Improve Workflow.Otol Neurotol. 2023 Aug 1;44(7):664-671. doi: 10.1097/MAO.0000000000003898. Epub 2023 Jun 6. Otol Neurotol. 2023. PMID: 37278159
-
A Systematic Review of Cochlear Implant-Related Magnetic Resonance Imaging Artifact: Implications for Clinical Imaging.Otol Neurotol. 2024 Mar 1;45(3):204-214. doi: 10.1097/MAO.0000000000004095. Epub 2024 Jan 17. Otol Neurotol. 2024. PMID: 38260952
-
Magnetic resonance perfusion for differentiating low-grade from high-grade gliomas at first presentation.Cochrane Database Syst Rev. 2018 Jan 22;1(1):CD011551. doi: 10.1002/14651858.CD011551.pub2. Cochrane Database Syst Rev. 2018. PMID: 29357120 Free PMC article.
-
Artifact Reduction 3.0 T MRI in Cochlear Implantation Recipient With Diametrically Bipolar Magnets.Laryngoscope. 2025 Aug;135(8):2901-2906. doi: 10.1002/lary.32114. Epub 2025 Apr 1. Laryngoscope. 2025. PMID: 40165611
-
Do Magnet®-accredited hospitals show improvements in nurse and patient outcomes compared to non-Magnet hospitals: a systematic review.JBI Database System Rev Implement Rep. 2015 Jul 17;13(6):168-219. doi: 10.11124/jbisrir-2015-2262. JBI Database System Rev Implement Rep. 2015. PMID: 26455752
References
-
- Nassiri AM, Sorkin DL, Carlson ML. Current estimates of cochlear implant utilization in the United States. Otol Neurotol 2022;43:e558–62.
-
- Smith-Bindman R, Kwan ML, Marlow EC, et al. Trends in use of medical imaging in US health care systems and in Ontario, Canada, 2000–2016. JAMA 2019;322:843–56.
-
- Philips B, Corthals P, De Raeve L, et al. Impact of newborn hearing screening: Comparing outcomes in pediatric cochlear implant users. Laryngoscope 2009;119:974–9.
-
- Fakurnejad S, Vail D, Song Y, Alyono J, Blevins NH. Trends in age of cochlear implant recipients, and the impact on perioperative complication rates. Otol Neurotol 2020;41:438–43.
-
- Cochlear. Getting an MRI with a cochlear implant: MRI guide. 2023. Available at: https://www.cochlear.com/us/en/home/ongoing-care-and-support/device-help... . Accessed September 12, 2024.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical