Simulating the Use of Discontinuous Patterned Hydrogel to Improve Inter-Electrode Resistance on Electrode Arrays
- PMID: 40485199
- DOI: 10.1111/aor.15030
Simulating the Use of Discontinuous Patterned Hydrogel to Improve Inter-Electrode Resistance on Electrode Arrays
Abstract
Background: A novel form of sensory stimulation aiming to treat spasticity has been developed, and a clinical trial is currently underway. This uses an electrode array controlled by a programmable 64-channel stimulator to spatially vary the electrical stimulation over time. However, when a continuous layer of hydrogel interfaces between the array and skin, stimulation spreads, causing lower current densities applied over larger areas of tissue. A new approach was developed, modeled, and tested, utilizing discontinuous patterned hydrogel to improve inter-electrode resistance on electrode arrays.
Methods: Finite-difference modeling was used to estimate stimulation distribution within the hydrogel and subcutaneous tissue under the electrode array. Repeated simulations modeled changes due to variations in hydrogel, skin, and subcutaneous tissue resistivity. Properties of both continuous sheets and patterned hydrogel were used for the simulation. Physical prototypes of the continuous and patterned hydrogel were manufactured and tested for comparison with the simulation.
Results: Simulation results showed a reduced spread of stimulation between electrodes when using the discontinuous patterned hydrogel compared to the continuous hydrogel. This was demonstrated consistently for all variations in hydrogel, skin, and subcutaneous tissue resistivity. Laboratory testing supported the simulation results and suggested the improved performance of the patterned hydrogel, compared with the continuous hydrogel, may become more substantial over time.
Conclusions: While the simulation only approximates the stimulation distribution on electrode arrays, the results do show potential benefits of utilizing discontinuous patterned hydrogel to increase inter-electrode resistance. Laboratory testing and initial feedback from the clinical trial support the results indicated in the simulations.
Keywords: electrical stimulation; electrode array; inter‐electrode resistance; patterned hydrogel; stimulation current density.
© 2025 The Author(s). Artificial Organs published by International Center for Artificial Organ and Transplantation (ICAOT) and Wiley Periodicals LLC.
References
-
- M. Slovak, J. Chindo, K. P. S. Nair, M. L. Reeves, B. Heller, and A. T. Barker, “Sensory Barrage Stimulation in the Treatment of Elbow Spasticity: A Crossover Double Blind Randomized Pilot Trial,” Neuromodulation 19 (2016): 220–226, https://doi.org/10.1111/ner.12383.
-
- A. Ali, A. D. McCarthy, M. Reeves, et al., “Sheffield Adaptive Patterned Electrical Stimulation (SHAPES) Therapy for Post Stroke Arm Spasticity: Study Protocol for a 3‐Arm, a Partially Blinded, Randomised Controlled Trial,” BMC Neurology 24 (2024): 437, https://doi.org/10.1186/s12883‐024‐03635‐x.
-
- B. W. Heller, A. J. Clarke, T. R. Good, et al., “Automated Setup of Functional Electrical Stimulation for Drop Foot Using a Novel 64 Channel Prototype Stimulator and Electrode Array: Results From a Gait‐Lab Based Study,” Medical Engineering & Physics 35, no. 1 (2013): 74–81, https://doi.org/10.1016/j.medengphy.2012.03.012.
-
- S. Agarwala, “Electrically Conducting Hydrogels for Health Care: Concept, Fabrication Methods, and Applications,” International Journal of Bioprinting 6, no. 2 (2020): 273, https://doi.org/10.18063/ijb.v6i2.273.
-
- C. Lim, Y. J. Hong, J. Jung, et al., “Tissue‐Like Skin‐Device Interface for Wearable Bioelectronics by Using Ultrasoft, Mass‐Permeable, and Low‐Impedance Hydrogels,” Science Advances 7 (2021): eabd3716, https://doi.org/10.1126/sciadv.abd3716.
Grants and funding
LinkOut - more resources
Full Text Sources